Checkout
2001989-351A
ApexPro™
7-11
6.
Read the current leakage indicated on DMM.
NOTE
If either reading is greater than the appropriate specification below, the
device under test fails. Contact GE Technical Support.
300 µA (0.3 volts on the DMM), and the device under test is powered from
100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is powered from a
centered-tapped 200-240 V/50-60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is powered from a
non-center-tapped, 200-240 V/50-60 Hz, single-phase circuit
7.
Set the
GND
switch on the leakage tester to
CLOSED
.
8.
Read the current leakage indicated on DMM.
9.
Set the polarity switch to
RVS
.
10. Read the current leakage indicated on DMM.
NOTE
If the reading is greater than the specification below, and the device under
test is powered from 100-240 V/50-60 Hz, the device under test fails.
Contact GE Technical Support.
100 µA (0.1 volts on the DMM), and the device under test is powered from
100-240 V/50-60 Hz
11. Set the power switch of the device under test to OFF.
Test completion
1.
Disconnect the leakage tester from the power outlet.
2.
Disconnect all test equipment from the device.
3.
Disconnect the device power cord from the leakage tester.
Transmitter checkout
Checkout procedure
NOTE
Due to IPX7 rating, the ApexPro CH and T14 Transmitters are not field
repairable.
1.
Perform visual inspection.
See Visual inspection on page 4-4.
2.
Perform the communication tests.
See Communications tests on page 7-17.
3.
If an oximeter is being used with the transmitter, perform the Oximeter
operational tests for the applicable oximeter.
Содержание ApexPro TT
Страница 9: ...2001989 351A ApexPro vii...
Страница 10: ...viii ApexPro 2001989 351A...
Страница 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Страница 20: ...1 10 ApexPro 2001989 351A Introduction...
Страница 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Страница 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Страница 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Страница 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Страница 84: ...4 22 ApexPro 2001989 351A Maintenance...
Страница 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Страница 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Страница 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Страница 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Страница 162: ...7 24 ApexPro 2001989 351A Checkout Unit Serial Number Institution Name Date Maintenance Repair Technician...
Страница 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Страница 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Страница 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Страница 191: ...2001989 351A ApexPro 1 1...