Checkout
2001989-351A
ApexPro™
7-23
Communication test
This procedure verifies that the telemetry system transmits and receives the data
correctly from the Accutracker.
1.
Attach the Accutracker to the ApexPro Telemetry system.
NOTE
You must have ECG leads with a shorting cable or simulator attached to the
ApexPro Telemetry system for this test to work.
CAUTION
Refer to the ApexPro operator’s manual for proper operation
guidelines and cuff/microphone placement.
2.
Place blood pressure cuff on the arm.
3.
Turn the Accutracker on and press the
button.
4.
When the display appears on the system screen, verify that the Accutracker
display numbers match the display numbers.
5.
Turn Accutracker off.
Repair log
Unit Serial Number:
Institution Name:
Date
Maintenance/Repair
Technician
Содержание ApexPro T14
Страница 9: ...2001989 351A ApexPro vii ...
Страница 10: ...viii ApexPro 2001989 351A ...
Страница 11: ...2001989 351A ApexPro 1 1 1 Introduction ...
Страница 20: ...1 10 ApexPro 2001989 351A Introduction ...
Страница 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview ...
Страница 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration ...
Страница 62: ...3 22 ApexPro 2001989 351A Installation and configuration ...
Страница 63: ...2001989 351A ApexPro 4 1 4 Maintenance ...
Страница 84: ...4 22 ApexPro 2001989 351A Maintenance ...
Страница 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting ...
Страница 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts ...
Страница 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module ...
Страница 139: ...2001989 351A ApexPro 7 1 7 Checkout ...
Страница 162: ...7 24 ApexPro 2001989 351A Checkout Unit Serial Number Institution Name Date Maintenance Repair Technician ...
Страница 163: ...2001989 351A ApexPro A 1 A Technical specifications ...
Страница 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility ...
Страница 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility ...
Страница 191: ...2001989 351A ApexPro 1 1 ...