Checkout
Ground (earth) integrity
Listed below are two methods for checking the ground (earth) integrity, “Ground
Continuity Test” and “Impedance of Protective Earth Connection.” These tests
determine whether the device's exposed metal and power inlet's earth (ground)
connection has a power ground fault condition.
Perform the test method below that is required by your Country/Local governing
safety organization.
Ground continuity test
Completion of this test is checked by the following steps:
1.
Disconnect the device under test from the power outlet.
2.
Connect the negative (-) lead of the DMM to the protective earth terminal
(ground pin in power inlet connector) or the protective earth pin in the Mains
plug (ground pin in power cord). Refer to the US 120 VAC power cord figure
above.
3.
Set the DMM to the milliohm (
mW
) range.
4.
Connect the positive (+) lead of the DMM to all exposed metal surfaces on the
device under test. If the metal surfaces are anodized or painted, scrape off a small
area in a inconspicuous place for the probe to make contact with the metal.
5.
Resistance must read:
0.1 ohm or less without power cord
0.2 ohms or less with power cord
Impedance of protective earth connection
This test, unlike a ground continuity test, will also stress the ground system by using
special ground bond testers.
This test normally is only required as a manufacturing production test to receive
safety agency compliance (i.e., IEC 60601-1).
Some country agencies do require this test after field equipment repairs (i.e.
Germany's DIN VDE 0751 standards).
Consult your country/local safety agency if in question.
Compliance is checked by the following steps:
1.
A current not less than 10A and not exceeding 25A from a current source with a
frequency of 50 or 60 Hz with a no-load voltage not exceeding 6V is passed for
at least 5s through the protective earth terminal or the protective earth pin in the
mains plug and each accessible metal part which could become live in case of
failure in basic insulation.
Ground
Pin
Содержание ApexPro T14
Страница 9: ...2001989 351A ApexPro vii ...
Страница 10: ...viii ApexPro 2001989 351A ...
Страница 11: ...2001989 351A ApexPro 1 1 1 Introduction ...
Страница 20: ...1 10 ApexPro 2001989 351A Introduction ...
Страница 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview ...
Страница 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration ...
Страница 62: ...3 22 ApexPro 2001989 351A Installation and configuration ...
Страница 63: ...2001989 351A ApexPro 4 1 4 Maintenance ...
Страница 84: ...4 22 ApexPro 2001989 351A Maintenance ...
Страница 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting ...
Страница 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts ...
Страница 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module ...
Страница 139: ...2001989 351A ApexPro 7 1 7 Checkout ...
Страница 162: ...7 24 ApexPro 2001989 351A Checkout Unit Serial Number Institution Name Date Maintenance Repair Technician ...
Страница 163: ...2001989 351A ApexPro A 1 A Technical specifications ...
Страница 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility ...
Страница 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility ...
Страница 191: ...2001989 351A ApexPro 1 1 ...