AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
GE HEALTHCARE
REV 8
28
6.
Repeat the above steps for each test point in Table 4: Test Points for AMX series
and record the
test results in Table 5: Results of the leakage current test under normal conditions for AMX series
for each test points.
7.
Repeat above steps with test conditions 2 & 3 as per Table 3: Test Condition for AMX series.
AMX Series
Condition
Power
Test Meter Polarity
1
ON
Normal
Table 3: Test Condition for AMX series
Test Point Components
Test Point Description
1
Collimator
Collimator Skin Spacers
2
Tube
X-ray Tube (On the HV Cable Cathode Ring)
3
Column
Rivets on the top column cover (Applicable only for AMX4+)
4
Bin
M4 Screw inserted into PEM nut on receptor bin
5
Tube Latch
Park Latch Assembly on the Chassis
Table 4: Test Points for AMX series
Note: Refer to Appendix 2 for the locations of the AMX series test points.
Expected Test Results
Test Point
Units Expected
Value
Measured
Value
Result
(Pass/Fail)
Collimator Skin Spacers
μA
≤ 3
00
X-ray Tube (On the HV Cable Cathode Ring)
μA
≤ 3
00
Rivets on the top column cover (Applicable
only for AMX4+)
μA
≤ 3
00
M4 Screw inserted into PEM nut on
receptor bin
μA
≤ 3
00
Park Latch Assembly on the Chassis
μA
≤ 3
00
Table 5: Results of the leakage current test under normal conditions for AMX series
TEST UNDER SINGLE FAULT CONDITION (SFC)
In SFC, a single fault means of protection against electrical hazard is defective or faulty.
To create Single Fault Condition, see Table 6: SFC Test conditions for 120VAC or 220VAC AC
Mains.
SFC Conditions
Condition
Power
Test Meter Polarity
X-Ray
1
On
Reversed
No
2
On
Earth Open
No
Table 6: SFC Test conditions for 120VAC or 220VAC AC Mains
To perform the Equipment Leakage Current Test under SFC, follow the steps listed below:
Note:
Refer to Table 6: SFC Test conditions for 120VAC or 220VAC AC Mains
for the test conditions
and Appendix 3: AC Power plug troubleshooting for the test equipment used. Refer to Table
4: Test Points for AMX series.
1.
Set up test condition 1 listed in Table 6: SFC Test conditions for 120VAC or 220VAC AC Mains, follow
step 2 below.