Absorber arms
Adjustable
Non-adjustable
Arm length: 30.5 cm/12 in
25.4 cm/10 in
Bag arm
height:
87 cm/34.3 in
91.5 cm/36 in
104 cm/40.9 in
Absorber
rotation:
85°
85°
Ventilator screen
Height:
7.6 cm/3 in
Width:
15.2 cm/6 in
Casters
Diameter:
12.5 cm/5 in
Brakes:
Single foot lever locks and
unlocks two front casters
Ventilator operating specifications
Ventilator (V
T
) parameter ranges
Tidal volume range:
20 to 1500 mL (Volume
Control and SIMV modes)
5 to 1500 mL (Pressure
Control Mode)
Incremental settings:
20 to 100 mL
(increments of 5 mL)
100 to 300 mL
(increments of 10 mL)
300 to 1000 mL
(increments of 25 mL)
1000 to 1500 mL
(increments of 50 mL)
Minute volume range:
0 to 99.9 L/min
Pressure (P
Inspired
) range:
5 to 60 cm H
2
O
(increments of 1 cm H
2
O)
Pressure (P
limit
) range:
12 to 100 cm H
2
O
(increments of 1 cm H
2
O)
Pressure (P
support
) range:
Off, 2 to 40 cm H
2
O
(increments of 1 cm H
2
O)
Rate:
4 to 100 breaths per
minute for Volume Control
and Pressure Control;
2 to 60 breaths per minute
for SIMV, PSVPro and
SIMV–PC+PSV (increments
of 1 breath per minute)
Inspiratory/expiratory ratio:
2:1 to 1:8 (increments of 0.5)
Inspiratory time:
0.2 to 5.0 seconds
(increments of 0.1 seconds)
(SIMV and PSV Pro)
Trigger window:
0 to 80% (increments of 5%)
Flow trigger:
0.2 to 1.0 L/min
(increments of 0.2 L/min)
1 to 10 L/min (increments
of 0.5 L/min)
Inspiration termination level:
5 to 75% (increments of 5%)
Backup mode delay:
10 to 30 seconds
(increments of 5 seconds)
Ventilation operating modes
Volume Control
Pressure Control
Synchronized Intermittent Mandatory Ventilation (SIMV)
Pressure Support (PSVPro) with Apnea Backup ventilation —
(optional)