29
Setting Auto Limits
Use the
Auto Limits
menu to
quickly set alarm ranges for ‘MV’,
‘TV’, and ‘EtCO
2
’ during mechanical
ventilation.
1.
Select the
Auto Limits
key.
The menu shows the current
measured values and the
proposed low and high alarm
limits. Check the proposed
parameters.
2.
Select
Confirm
to use the
proposed low and high alarm
limits.
3.
Select
Cancel
to leave the alarm
limits unchanged.
4.
Select
Case Default Limits
to
set the alarm limits to the case
default limits.
Viewing Trends
Use the
Trends
menu to view
patient trends and set the time
scale.
1.
Select the
Trends
key.
2.
Select the desired view.
3.
Select
Scroll
to move through
the current trend view.
4.
Select
Time Scale
to select
the desired scale from the
dropdown menu.
5.
Select
Next Page
to view
additional parameters.
6.
Select
Close
to close the menu.
1
1
3
4
5
6
2
3
2
4