28
Additional alarm settings
The following alarms can be turned off to reduce
nuisance alarms during manual ventilation if clinically
appropriate:
1. CO
2
Alarms:
Use the
CO
2
Alarms
setting to turn off
the CO
2
‘Apnea’ alarm,
EtCO
2
low, EtCO
2
high,
and
FiCO
2
high
and
Absorbent OK?
alarms during
manual ventilation.
2. Vol Apnea Alarm
: Use the
Vol Apnea Alarm
setting
to turn off the volume apnea alarm during manual
ventilation.
3. MV/TV Alarms
: Use the
MV/TV Alarms
setting to
turn off the MV and TV alarms. If
MV/TV Alarms
is
set to
Off
during manual ventilation, the alarms
remain off until the Bag/Vent switch is set to Vent or
the
MV/TV Alarms
is set to
On
. If the
MV/TV
Alarms
is set to
Off
during mechanical ventilation,
the alarms remain off when manual ventilation
starts.
Note
:
CO
2
Alarms
and
Apnea Alarm
selection will not
be visible unless enabled by a Super User.
CO
2
Alarms
and
Vol Apnea Alarm
can only be turned off while in
manual ventilation.
1
2
3