GE
LOGIQ F S
ERIES
D
IRECTION
5446617-100, R
EVISION
10
B
ASIC
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4 - 13
Width
Yes
You can widen or narrow the size of the sector angle to
maximize the image's region of interest (ROI).
Tilt
Yes
You can steer the sector angle to get more information
without moving the probe while in B-Mode, M-Mode,
Doppler Mode, and Color Flow Mode.
Reverse
No
Flips the image 180 degrees left/right.
Dynamic Range
No
Dynamic Range controls how echo intensities are
converted to shades of gray, thereby increasing the
adjustable range of contrast.
Line Density
Yes
Optimizes B-Mode frame rate or spatial resolution for the
best possible image.
Map
No
The system supplies B, M, and Doppler Mode system
maps.
Frame Average
No
Temporal filter that averages frames together, thereby
using more pixels to make up one image. This has the
effect of presenting a smoother, softer image.
Colorize
No
Colorize is the colorization of a conventional B-Mode
image or Doppler Spectrum to enhance the user's ability
to discern B, M, and Doppler Mode intensity valuations.
Colorize is NOT a Doppler Mode.
NOTE: You can colorize realtime or CINE images or
Timeline CINE.
Colorizes the gray scale image to enhance the eye's
discrimination capability. Spectrum Colorize colorizes the
spectrum as a function of power using the inverse of the
Colorize map for the signal intensity in each Doppler line.
Colorize enhances the visibility of the spectrum's
characteristics and enhances your ability to identify
spectral broadening and the edge contours of the
spectrum used to define the peak frequency/velocity.
The colorize bar displays while Colorize is activated.
Edge Enhance
No
Edge Enhance brings out subtle tissue differences and
boundaries by enhancing the gray scale differences
corresponding to the edges of structures.
Rotation
No
You can flip the image up/down.
CAUTION:
When reading an rotated image, be careful to
observe the probe orientation to avoid possible confusion
over scan direction or left/right image reversal.
Rejection
No
Selects a level below which echoes will not be amplified
(an echo must have a certain minimum amplitude before it
will be processed).
Suppression
No
Suppresses the noise in the image.
Table 4-2
B Mode Controls
Control
Possible Bioeffect
Description/Benefit
Содержание 5478035
Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10 ...
Страница 2: ......
Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions ...
Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii ...
Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii ...
Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv ...
Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v ...
Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi ...
Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii ...
Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii ...
Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix ...
Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x ...
Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii ...
Страница 167: ... 2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...