Page 28
March 2022
7.4.2
Space required
To store the 1.5T or 3.0T HD 8ch Wrist Array coil and baseplate, a storage space of 55cm cm/21.5 in
(width) x 45 cm/17.6 in (depth) x 35cm/13.7 in (height) is required.
Disassembly of Coil
7.5.1
Unlatch Coil
Unlatch the coil by pushing the lock toward the cable end of the coil
and pulling the coil open (up).
7.5.2
Detaching Coil
Slide the coil out of the round cutouts.
Product End of Life
The functional life and proper performance of coils are guaranteed only while the coil is being used on
the MR system (hardware/software level) specified at the time of purchase. Upgrades or other
modifications to the system software and/or hardware may affect compatibility.
During ongoing use of the coils,
immediately discontinue use of this device if the following symptoms of
age appear:
Frayed Cable
- If the insulation or housing is broken, or if the cable is frayed, immediately discontinue use
of the device.
Damaged connector and pins
- damaged cable pins and connector may damage the connector and affect
coil or system performance.
Cracked coil casing
–
cracked coil casings can expose electrical components, rendering the coil unsafe for
the patient and/or user.