GE Druck PV 411A Скачать руководство пользователя страница 6

GE Druck PV 411A User Manual

K258 Issue No. 3

2

    

Operation

WARNINGS

1

B

EFORE

 

APPLYING

 

PRESSURE

MAKE

 

SURE

 

ALL

 

CONNECTIONS

 

ARE

 

CORRECT

 

AND

 

EQUIPMENT

 

IS

 

INTERNALLY

 

CLEAN

 

AND

 

FREE

 

FROM

 

DAMAGE

.

2

M

AKE

 

SURE

 

THAT

 

ALL

 

EQUIPMENT

 

IS

 

TO

 

THE

 

CORRECT

 

PRESSURE

 

RATING

.

3

D

O

 

NOT

 

EXCEED

 

THE

 

MAXIMUM

 

OPERATING

 

PRESSURE

 

STATED

 

IN

 

THE

 

SPECIFICATION

.

4

O

BSERVE

 

THE

 

RELEVANT

 

HEALTH

 

AND

 

SAFETY

 

PRECAUTIONS

.

Note:

Wherever possible, use o-ring seals in the connection ports this is the 
recommended method of sealing.

Key to 

1 Inlet (reservoir) port
2 Outlet  port
3 Volume adjuster
4 Limit adjuster
5 Scissor-action handles
6 Outlet  port
7 Selector  valve
8 Pressure relief valve (see figure 3)

Pressure/vacuum Configuration 

In pressure mode, air/fluid is drawn in through the inlet port (1) on the top of 
the pump and forced out through the two outlet ports (2 & 6).
In vacuum mode, the air/fluid flow is reversed as air/fluid is drawn in through 
the top and rear outlet ports (2 & 6) and expelled through the inlet port (1).

Содержание PV 411A

Страница 1: ...PV 411A K258 gGE Druck ...

Страница 2: ...3 4 1 2 2 3 1 2 3 3 7 1 2 6 5 4 1 8 3 ...

Страница 3: ...ted in the specification This manual contains safety and operating instructions that must be followed to make sure of safe operation and to keep the equipment in a safe condition The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage Use suitably qualified personnel and good engineering practice for all procedures in this manual P...

Страница 4: ... 5 Hydraulic Operation 6 Fluid reservoir 6 Fitting 6 Filling 7 Priming the system 7 Vacuum priming 7 Pre filling 7 Generating hydraulic pressure 8 Setting pressure relief valve 8 Fault finding 9 Annex A Returned Goods Procedure 10 Approved service agents inside rear cover C degrees Celsius mbar millibar BSP British Standard Pipe thread mm millimetre cm centimetre mmH2O millimetres of water cSt cen...

Страница 5: ... priming process The volume adjuster 3 generates the required system pressure A fluid reservoir see Figure 2 screwed into the inlet port 1 marked RESERVOIR on the top of the hand pump completes the hydraulic circuit A pressure relief valve 8 can be adjusted to set a pressure between 30 and 700 bar Specification Pressure range Pneumatic 0 to 60 bar Vacuum 0 to 0 95 bar Hydraulic 0 to 700 bar Maximu...

Страница 6: ...herever possible use o ring seals in the connection ports this is the recommended method of sealing Key to 1 Inlet reservoir port 2 Outlet port 3 Volume adjuster 4 Limit adjuster 5 Scissor action handles 6 Outlet port 7 Selector valve 8 Pressure relief valve see figure 3 Pressure vacuum Configuration In pressure mode air fluid is drawn in through the inlet port 1 on the top of the pump and forced ...

Страница 7: ...timum viewing angle and bench top operation Placing the pump on the bench allows greater force to be applied to the scissor action handles Internal stops prevent further rotation of the pump head Note Do not use excessive force when rotating the pump head Selector Valve Pressure Turn the selector valve fully clockwise in position Vacuum Turn the selector valve fully anti clockwise out position Ven...

Страница 8: ...lockwise out to decrease the pressure Vacuum Turn the selector valve fully anti clockwise out position Turn the volume adjuster to the mid position In this position fine adjustments of the generated vacuum can be made Using the scissor action handles generate the approximate vacuum then turn the volume adjuster clockwise in to decrease the vacuum or turn the volume adjuster anti clockwise out to i...

Страница 9: ...osphere by slowly turning anti clockwise the selector valve 7 to the centre position Vacuum 1 Turn the selector valve 7 fully anti clockwise out position 2 Operate the scissor action handles 5 to generate the approximate vacuum allow time for thermal stabilization 3 If necessary use the volume adjuster 3 to adjust the required vacuum 4 After attaining the required vacuum operate the scissor action...

Страница 10: ...acrylic providing a clear view of the contents The reservoir can be removed from the pump without the need to empty the fluid a self sealing connection prevents leakage The spring loaded reservoir cover 2 seals under atmospheric pressure conditions but vents excess internal pressure harmlessly in the event of inadvertent pressurization The reservoir connects to the inlet port on the top of the pum...

Страница 11: ...ty on the pump Note This method should not be used with vacuum sensitive equipment Turn selector valve fully out anti clockwise vacuum position Turn the volume adjuster valve fully anti clockwise out position Pump until bubbles stop appearing in the reservoir this generates a vacuum in the system Turn the selector valve fully in this action releases the vacuum and rapidly fills the system with flu...

Страница 12: ...out anti clockwise and turn the selector valve 7 to depressurize the pump Disconnect from the pipes and equipment Setting Pressure Relief Valve The pressure relief valve PRV can be set at pressures from 30 to 700 bar CAUTION DO NOT EXCEED 700 BAR THIS CAN DAMAGE THE INTERNAL SEALS OF THE PRV AND THE PV411A If the system pressure exceeds the set pressure the PRV opens and vents fluid through the in...

Страница 13: ...lector valve 7 is in the correct position and properly tightened 2 Allow sufficient time after pressure generation for the temperature to stabilize The larger the system volume the longer the time for thermal stability 3 Check the adaptors flexible pipe and connections between the pump and equipment under test for leaks Tighten any loose joints and replace any seals that are worn or damaged 4 In h...

Страница 14: ...ation RGA number providing the following information Product i e PV 411A Pressure medium i e pneumatic hydraulic Serial number Details of defect work to be undertaken Operating conditions Safety Precautions You must also tell us if the product has been in contact with anything hazardous or toxic and the relevant COSHH references and precautions to be taken when handling Important notice Service by...

Отзывы: