Revision 1.1 

 

GM ECx Instruction Manual 

 

11

3.5.1 Antenna 

Selection 

GDS Corp offers locally mounted dipole antennas suitable for either Division 1 or Division 2 
classified areas

 as shown in Figure 3-3 below.  Antennas may also be remotely mounted if it is 

necessary to raise the height to achieve better RF L.O.S. (Line of Sight). 
 

 

 

Figure 3-3: Local Antennas for Hazardous Locations  

 

 

3.6 

Sensor Installation (with Smart / Simple Sensor Definition) 

Sensors used with the 10-0247 sensor head (see Figure 3-5) are considered “smart” since they 
include a memory device that uploads sensor information to the GM ECx anytime a new sensor is 
installed.  The GM ECx may also accommodate other industry standard “simple” electrochemical 
toxic / oxygen sensors, without the memory device, by accepting their wiring into TB2 of the 10-
0232 Display Assembly (see Figure 3-4).  The 10-0247’s 

Smart Sensor

 cable connects to S1 of 

the 10-0232 Display Assembly (see Figure 3-4).  This unique 

Smart Sensor Interface

 may be 

used to configure smart sensors and / or GM ECX’s from a PC rather than entering all variables 
via the magnetic keypad. 

Sensors used with the smart sensor head (see Figure 3-5) are considered “smart” since they 
include a memory device that uploads sensor information to the GM ECx anytime a new sensor is 
installed.  The GM ECx may also accommodate other industry standard “simple” electrochemical 
toxic / oxygen sensors, without the memory device, by accepting their wiring into TB2 of the 
Display Assembly (see Figure 3-4).  Sensor heads with the 8-wire 

Smart Sensor

 cable connects 

to S2 of the 10-0291 Radio Assembly (see Figure 3-4).  This unique 

Smart Sensor Interface

 

may be used to configure smart sensors and / or GM ECx from a PC rather than entering all 
variables via the magnetic keypad. 

Содержание GASMAX ECx

Страница 1: ...mart Sensor Interface Important Read and understand contents of this instruction manual prior to use Improper use of equipment could result in instrument malfunction or serious injury GDS Corp 2513 Hwy 646 Santa Fe Texas 77510 409 927 2980 409 927 4180 FAX www gdscorp com tech gdscorp com ...

Страница 2: ...and GM ECX Signal Conditioning 13 SECTION 4 INITIAL START UP 13 4 1 Transmitter Configuration Menu 13 4 2 Initial Toxic Oxygen Monitor Start Up 13 4 2 1 Initial Toxic Oxygen Monitor Span Check 14 SECTION 5 OPERATING INSTRUCTIONS 14 5 1 CAL MODE Routine Sensor Calibrations 14 5 2 ALARM OPERATION 16 5 2 1 ALARM 3 UNDERSTANDING FAULT LEVEL OPERATION 16 SECTION 6 SETUP MENU CONFIGURATION 17 6 1 Menus ...

Страница 3: ...Cx Periodically test for correct operation of the system s alarm events by exposing the monitor to a targeted gas concentration above the High Alarm setpoint Do not expose the GM ECx to electrical shock or continuous severe mechanical shock Protect the GM ECx from dripping liquids and high power sprays Use only for applications described within this manual CAUTION FOR SAFETY REASONS THIS EQUIPMENT...

Страница 4: ...ature allows event logging of calibrations and alarm trips for review on the LCD readout Compatible sensors include an 8 wire Smart Sensor interface capable of configuration data uploads to the monitor as well as traditional Simple sensors with three wire interface A separate PC compatible USB Interface device part 10 0252 allows Smart sensors to be preloaded with configuration variables via a PC ...

Страница 5: ...l 0 32 and System ID 0 255 to communicate see section 6 7 There should never be two servers with the same Hop Channel System ID settings in the same coverage area The Server must be in a powered location and should be centrally located since all Clients must receive the server s beacon in order to communicate Correct planning and design of wireless systems are imperative for ensuring a successful ...

Страница 6: ... the radio to transmit Five minutes or fifty sequential reads when there is no A1 or A2 alarm Any receiving controller will report a Communications Error if the GM ECx does not reply for periods of greater than 18 minutes A3 and FAIL alarms do not increase radio transmission rates Every six seconds if there is an A1 or A2 level alarm Upon entry into CAL mode a value of 75 15 6 scale is transmitted...

Страница 7: ...0291 The 10 0291 RF module mounts piggy back to the back of the Display assembly as shown in Figure 2 3 The MMCX RF connector connects to the antenna fitting s pigtail coax cable There is also a slender 4 conductor cable between the 10 0291 and the 10 0299 battery I O PCB bolted to the bottom of the enclosure ...

Страница 8: ...there are no significant currents Lighter than air gases should have the monitors mounted 12 18 inches 30 45 centimeters above the potential gas leak and heavier than air gases should be this distance below Even though the GM ECx is designed for rugged service sensors should be protected from environmental damage from water snow shock vibration and dirt 3 3 Mounting the Enclosure The GM ECx standa...

Страница 9: ...ounting flanges WARNING Qualified personnel should perform the installation according to applicable electrical codes regulations and safety standards Insure correct cabling and sealing fitting practices are implemented Do not aim the sensor pointing upward Install the GM ECx to a wall or bracket using the predrilled mounting flanges with I D 0 25 on 5 5 inch centers Figure 3 1 If conduit is rigid ...

Страница 10: ...sses in the system For best RF line of site the combined height of both antennas must exceed the Fresnel zone diameter see below Dist between ant s Fresnel zone diameter Freespace loss dB 1000 ft 300 m 16 ft 4 9 m 81 1 Mile 1 6 km 32 ft 9 7 m 96 5 miles 8 km 68 ft 20 7 m 110 10 miles 16 km 95 ft 29 m 116 Example The RF900 radio modem has the following parameters Maximum RF900 TX power setting 30 d...

Страница 11: ...wiring into TB2 of the 10 0232 Display Assembly see Figure 3 4 The 10 0247 s Smart Sensor cable connects to S1 of the 10 0232 Display Assembly see Figure 3 4 This unique Smart Sensor Interface may be used to configure smart sensors and or GM ECX s from a PC rather than entering all variables via the magnetic keypad Sensors used with the smart sensor head see Figure 3 5 are considered smart since t...

Страница 12: ...is shown after power up upon installation of a new smart sensor or by viewing INPUT type in the SENSOR SETTINGS INFO menu section 5 5 SMART SENSOR Type Cat Bead Span 100 Zero 0 SN xxxxxx Born On 01 28 04 Last Cal 04 05 04 ANY key to Exit ERROR CODE 01 Incorrect Sensor installed Install correct sensor or update transmitter SEE MANUAL EDIT key to update Any other to abort If installed sensor type do...

Страница 13: ...ECTION 4 INITIAL START UP 4 1 Transmitter Configuration Menu Figure 4 1shows the GM ECX XMITTER CONFIG menu used to activate channels precisely calibrate 4 20mA outputs and set time date Its menus are set at the factory and typically not needed by the user To access from any data display press and hold the NEXT key for 5 seconds until the screen appears requesting a special key sequence 4 UP keyst...

Страница 14: ...it output indicates CAL MODE by transmitting 75 counts It then transmits 200 counts during the subsequent CAL PURGE delay to prevent external alarms during calibration CAL MODE automatically exits if no keystrokes are detected after 5 minutes The GM ECx wireless 10 bit output range is 200 to 1000 counts for 0 to 100 of full scale The GM ECx indicates CAL MODE by transmitting 75 counts 15 6 FS to r...

Страница 15: ...calibration WARNING The SPAN gas used must match the value specified since this is what the GM ECX will indicate after a successful SPAN calibration The Cal Span Value may be edited if it becomes necessary to apply a different gas concentration see Cal Span Value in section 3 4 5 If the SPAN calibration is successful the display flashes REMOVE CAL GAS and starts the CAL PURGE delay see section 6 6...

Страница 16: ...ffset Gain constants Measurement Name LEL 2 Measurement Name LEL 0 PGA xxxx Figure 5 2 Cal Mode Flow Chart and Menus 5 2 ALARM OPERATION GM ECX s have front panel LED indicators for Alarm 1 Alarm 2 and FAIL CAUTION GM ECX Alarm LED indicators function even without the presence of the 10 0234 Relay option With battery power operation to conserve power alarm LED s only flash during alarm events 5 2 ...

Страница 17: ...each unit The Educator allows configuring up to six GM ECx s at the PC and transporting their configuration data to each location This is useful when numerous installed units must be configured The GM ECx s configuration menus are shown in Figure 6 1 Figure 6 1 Configuration Menu Tree 6 2 Configuration Using the Magnetic Wand Passing the magnetic wand past the EDIT key from either data display dis...

Страница 18: ... to be displayed when 20mA 100 is the GM ECX output The highest reading allowed is 9999 includes negative polarity sign and one decimal point Polarity is only indicated for negative readings Decimal Points sets the resolution of the LCD readings and may be for 0 1 or 2 Example ZERO readings for 0 1 2 DP s respectively are 0 0 0 0 00 Cal Span Value sets what upscale value must be applied when perfo...

Страница 19: ...ge in this direction A3 has a default negative 5 of range Set Point with Low Trip set for YES This makes it function as a FAULT alarm and trip when the monitored value is more than 5 out of range Dead Band has a minimum value of 1 and a maximum value of 10 It is useful for preventing alarm cycling when the monitored value is hovering around the set point EXAMPLE With a range of 0 100 ppm if Dead B...

Страница 20: ...u is therefore not available with a smart sensor connected Recall Cal History recalls each successful calibration These dates may be reviewed by scrolling with the UP DOWN keys Input Type indicates what kind of input or sensor the GM ECX is configured to accept and is typically pre configured at the factory There are three Input Type possibilities consisting of simple EC negative simple EC positiv...

Страница 21: ...e RTU address for this particular GM ECx Controller channels receiving this monitor s data must also be configured with this matching Remote ID address System ID may be set from 1 255 and is similar to a password character or network number and makes network eavesdropping more difficult A transceiver will not go In Range of or communicate with a transceiver operating on a different System ID TX Po...

Страница 22: ...ineering unit range should be 0 5 00 VDC Even if this channel reads VALUE from the GM ECx the battery voltage is displayed at the bottom of this C1 menu as shown in Figure 6 8 Figure 6 8 C1 Base Station Data From Menu 6 8 SYSTEM SECURITY The SYSTEM SECURITY menu offers two levels of protection A LOW level allows CAL MODE sensor calibrations but requires the 4 digit Pass Code prior to altering menu...

Страница 23: ...these menus Back up the current configuration prior to altering any Technical menus in case Restore is required later see section 6 3 The TECHNICIAN ONLY menu group in Figure 7 1 contains items that are factory configured depending upon the type sensor and input connected to the GM ECX They should not be tampered with after installation If configured incorrectly some items will prevent monitoring ...

Страница 24: ...ve range value of 0 1023 CAUTION For standard installations this is a factory adjustment Do not use the PreAmp Gain Set menu for calibrating sensors It should only be adjusted if a new measurement gas or input range is required PGA Setup Gain 25 0 Counts 1011 Reading 100 Up Dn to Change Next to exit Figure 7 3 PreAmp Gain Adjust PGA Menu 7 4 Simple Sensor Input Type Technicians only Smart sensors ...

Страница 25: ...y The Zero Cal Value menu entry allows the zero calibration value to be set for something other than a zero reading In rare cases it may be necessary to perform Zero calibrations at some other engineering unit reading than zero Do not exceed 25 of full scale 7 6 Raw Min Max Counts Technicians only The Raw Min Max Counts menus determine the range of 10 bit analog to digital A D converter counts tha...

Отзывы: