10/96
EN
Any devices used with TU must be compatible with the gas for which TU
is intended for.
Always make sure the device is connected first and then used on the
patient.
7.2.2. CONNECTION OF MEDICAL DEVICES TO QUICK CONNEC-
TORS
•
Connect the medical devices used in accordance with EN ISO 9170-1
and the appropriate national standards SS8752430.
•
Check whether there is gas on the outlet of the medical device con-
nected (e.g. by setting the flow rate of the flow rate meter connected).
•
Use the medical device connected in accordance with the manufac-
turer’s instructions.
7.3. AFTER EVERY USE
Disconnect the medical devices used from the terminal unit.
8. CLEANING
Remove dirt with a soft cloth damped in oil free oxygen compatible soap
water and rinsed with clean water. Disinfection can be carried out with an
alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and
that they are compatible with brass, plastic materials of components, and
gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature.
9. LIFETIME OF THE PRODUCT, MAINTENANCE
AND SERVICING
9.1. BATCH NUMBER AND PRODUCTION DATE
The number stamped on the terminal unit body consists of the following
data: GCE logo XXXXXXX
XXXXXXX: batch number
For example: The number 7366506 indicates a product with the batch
number 7366506.
9.1.1. PRODUCT LIFETIME AND WASTE MANAGEMENT
Housing - the maximum lifetime of this part of the product is given by the
lifetime of non-metallic materials used in the device and equals to the life-
time of materials used in the pipeline distribution system.