20
HiToP
®
Metabol
gbo Medizintechnik AG
Version 1.4
8
Technical data
Mains voltage and frequency:
115, 230 V, 48 - 62 Hz
Current consumption:
With 115 V: max. 700 mA
With 230 V: max. 350 mA
Mains fuses:
With 115 V: 2,5 AT
with 230 V: 2,5 AT
Output current:
max. 250 mA
Output voltage:
max. 70 V eff.
Admissible loaded
impedance:
30
...5 k
MDD device class:
IIa
Safety class:
I in accordance to IEC 601 / VDE 0750
Safety degree:
BF in accordance to IEC 601
Protection against ingress of
water:
IP X1
Dimensions:
6 cm x 23 cm x 32 cm (H x D x W)
Weight:
max. 3,7 kg (without accessories)
Color:
white RAL 9002 and grey RAL 7016
Display:
LCD backlighted, 122 x 32 dots full graphic
Buffer battery:
CR 2032
Surrounding conditions:
Operation
of the
device:
Temperature range +10 °C ... +40 °C
Relative air humidity 30 ... 75 %
Transport
and storage:
Temperature range +5 °C ... +50 °C
Relative air humidity < 90 %, none
condensing
Types of current:
Sinusoidal currents of 4096 - 32768 Hz
gbo Medizintechnik AG reserves the right to modify the design and specification
without prior notice.