3.3 Temperature Settings
The air temperatures are identified on the
operator control panel and indicate the average
air temperature at the hose end using a Quilt.
NOTE: The air temperature around the patient
is affected both by the ambient room
temperature and the use of an insulating
blanket on top of the Quilt.
3.4 Safety System
Dual Patient Safety Temperature Limit
Thermostats
Either of two independently operating
thermostats will shut off the TC3001 at a
preset high limit temperature. The heater
element and blower will remain off until the
thermostat is manually reset.
HeaterOvertemp
Heater overtemp thermostat will shut off the
TC3001 in the event of a blower failure or
lack of air movement. The heater element
and blower will remain off until the
thermostat is manually reset.
High Temp Indicator
The
REMOVE FROM USE
indicator will light
on the front panel when a patient safety
thermostat or heater overtemp thermostat
has tripped. Also, the heating element and
blower will turn off.
HEPA Filter Interlock Switch
The
REMOVE FROM USE
indicator will light
on the front panel when the HEPA filter has
been installed incorrectly or is missing. Also,
the heating element and blower will turn off.
3.5 Quilts
All Quilts are made of a nonwoven layer bonded
to a plastic film.
Quilt material meets U. S. flammability standards:
•
NFPA 702 “Normal Flammability”
•
CPSC Part 1632
•
California Title 19, Subchapter 7
Clear drape material meets NFPA 702 “Normal
Flammability.”
Section 3 - Specifications
3.0 Specifications
3.1 Physical
Dimensions
42 cm x 28 cm x 27 cm
Weight
6.8 kg
Enclosure
Thermoplastic
Filter
HEPA filtration to 0.2 micron
particle size
3.2 Electrical
Input
100 V (+10%, –5%),
50/60 Hz, 14 amps max
Motor
1/25 HP, single phase
Heater
1200W heating element
Power Cord
4.6 meters, 14 AWG cord with
hospital grade plug.
Circuit Breaker
15 amp
Current Leakage (Earth)
100 microamps max.
Ground Resistance
0.15 ohms nominal;
0.5 ohms maximum
Ambient Operating Temperature
16°C to 29°C
4
Table 1 - Air Temperatures
Temperature Setting
Temperature
Fan only
32°C
38°C
43°C
46°C
Содержание Thermacare TC3001
Страница 1: ...Service Manual Thermacare TC3001 Convective Warming Unit P N 11897 000 02 06...
Страница 30: ...28 Section 13 Wiring Diagram Figure 13 Wiring Diagram ATTENTION Static Sensitive Devices...
Страница 31: ...29 Section 13 Wiring Diagram cont d...
Страница 32: ...GAYMAR INDUSTRIES INC 10 Centre Drive Orchard Park NY 14127 2295 USA Phone 716 662 8636 FAX 716 662 0730...