4. Repeat step 2 for
43°C
setting. Average
temperatures and record.
Allowable 43°C temperature:
41.7°C to 45.0°C average
5. Repeat step 2 for
46°C
setting.
Allowable 46°C temperature:
44.7°C to 47.5°C average
6. If the temperatures in steps 2 through 5 are
not correct, do not put the Power Unit into
service. Contact your dealer for assistance.
7. Unplug the Power Unit.
7.8 Ground Resistance
Check grounding resistance:
1. Use a current leakage/ground resistance
tester to measure the resistance between the
grounding pin on the power entry module
and the I. V. pole clamp. An unplated area on
the underside of the I. V. pole clamp has
been provided for this test. Record the value.
2. The value should be less than 0.5 ohms.
7.9 Current Leakage
Measure current leakage:
1. Measure and record the maximum earth
current leakage (ground open). An unplated
area on the underside of the I. V. pole clamp
has been provided if needed for this test.
Measure at all combinations of:
•
line polarity
•
neutral open/closed
2. The current leakage should not exceed 100
microamperes in any condition.
Section 7 - Functional Check and Safety Inspection (cont’d)
15
7.10 Completing the Functional Check
and Safety Inspection
If the Power Unit has passed the
FUNCTIONAL
CHECK AND SAFETY INSPECTION
for all
requirements of procedures 7.1 through 7.9, the
Power Unit should be considered operational and
suitable for return to service.
This completes the recommended
FUNCTIONAL
CHECK AND SAFETY INSPECTION
for the TC3001
series Power Unit.
Содержание Thermacare TC3001
Страница 1: ...Service Manual Thermacare TC3001 Convective Warming Unit P N 11897 000 02 06...
Страница 30: ...28 Section 13 Wiring Diagram Figure 13 Wiring Diagram ATTENTION Static Sensitive Devices...
Страница 31: ...29 Section 13 Wiring Diagram cont d...
Страница 32: ...GAYMAR INDUSTRIES INC 10 Centre Drive Orchard Park NY 14127 2295 USA Phone 716 662 8636 FAX 716 662 0730...