Troubleshooting | 53
6. TROUBLESHOOTING
6.1 GENERAL TROUBLESHOOTING
Symptom
Possible Cause
Solution
Control box not circulating blood
dispensing bag fluid
Power is not being delivered to the
control box
Check that the port on the
control box that the power
cable plugs into is not
damaged. Remove the power
cable from the control box and
inspect the port. The metal pin
inside the port should not be
damaged or bent in any way.
Disconnect and firmly reconnect
the power cable that plugs into
the port on the control box. It
may be that the power cable
was not properly connected.
Check that the outlet the
power cable is plugged
into is a working one.
There is no fluid flowing through
the vasculature of the arm when
the system is turned ON; no fluid is
exiting the drainage tubings.
The fluid flow is restricted
Ensure that all of the tubing is
unclamped.
Disconnect and clean the
system. It is possible that
leftover artificial blood
concentrate in the system
clogged the vasculature.
Reconnect the system and try
again.
Roll down the arm skin and
inspect the veins. Ensure that
the veins are not kinked. Tape
down veins to keep them in
place if necessary. Replace veins
as needed.
The arm is leaking from the puncture
marks on the arm skin
The arm skin and/or veins have been
punctured with too large a needle or
too many times
Replace the arm skin and veins
as needed. Refer to section "5.
Routine Maintenance" on page
38