ORIUS™ Use
r
’s Guide Series Rev 3
2-1
2 Getting Started
The following steps will enable the first-time user to switch on the digital camera, launch DigitalMicrograph and
prepare for the acquisition of an image.
•
Make sure that the viewing chamber is at the proper vacuum.
•
Make sure the water and the pressure lines are correctly connected to
the camera and proper water pressure is maintained.
•
Switch on the CCD camera controller.
•
Wait for the lights on the front of the controller to change to the first
three on, the rest off. (Power, Ready and Camera lit green)
•
Turn the computer on (if it is not on yet) and double click in the
DigitalMicrograph icon to launch the application.
•
Set the camera to cool by picking the option “Temperature” under the
camera menu. Next, select a temperature of 10C. The camera
temperature will stabilize in about 2-5 minutes, then the camera
controller cooler indicator light will go from yellow to green.
2.1 The Workplace
When DigitalMicrograph is launched, you will see the commonly used tools on the Toolbar at the top of the
main window, the Image Viewing and Acquisition tools (right side), and the Result window as shown in Figure
2-1. The windows seen are composed as “floating” windows which are dockable to allow customization.
Содержание ORIUS SC 200D
Страница 40: ...ORIUS 833 User s Guide Series Rev 3 IV Figure A4 Camera Pneumatic Interface Details...
Страница 41: ...ORIUS 833 User s Guide Series Rev 3 V Figure A5 Water lines...
Страница 43: ...ORIUS 833 User s Guide Series Rev 3 VII Appendix 2 Environmental specifications and safety requirements...
Страница 45: ...ORIUS User s Guide Series Rev 3 IX Appendix 3 Camera specifications...
Страница 47: ...ORIUS 833 User s Guide Series Rev 3 XI Appendix 4 CE certification...
Страница 48: ...ORIUS 833 User s Guide Series Rev 3 XII...