WARNINGS
•
Do not inhale output gas
•
Do not immerse in water
•
Ensure dry, compatible materials in delivery system
•
Allow sufficient time for gas concentration to stabilise after switch on
•
Always purge the instrument after use
•
Under extreme fault conditions the unit could generate up to 100ppm
Chlorine
TROUBLESHOOTING
Q
The ‘CELL LIFE REMAINING’ indicator shows 0?
A
The generating cell can be replaced by either:
a. returning the GazCalTM to Gastech (or local representative) or
b. fitting a new cell yourself – new cells from Gastech come with full in-
structions.
Q
Low battery symbol appears on the display?
A
Charge instrument with ‘GazCalTM CHARGER’ for eight hours
Q
The ‘BLOCKAGE’ LED is illuminated?
A
1. Check that the gas outlet pipe is no longer than 1m in length,
as supplied by Gastech.
2. Ensure that any adaptors on calibration cups fitted to the outlet do not
cause restrictions to the gas flow.
3. Check gas outlet pipe for blockage.
4. Replace internal air filter (as detailed on page 4).
Q
Your GazCalTM is dropped or immersed in liquid?
A
Return the instrument to Gastech or the local representative for checking.