INTRODUCTION

The GazCalTM is a gas generator that produces concentrations of Chlorine 
(CL2) for testing and calibrating gas detectors. The unit is battery operated and 
is designed for field applications.

The GazCalTM uses an electrochemical cell to produce the gas. A cloudy 
controlled airflow, provided by an internal pump, then dilutes the gas to the 
required concentration.

SPECIFICATION

Concentration range:

Adjustable between 0.5 and 20ppm ±10% or 0.25ppm 
whichever is greater

Calibration:

0.5 - 5ppm or 5 - 20ppm

Flow rate:

0.5 l/min or 1 l/min

Cell Life:

500ppm hours

Power Supply:

Internal rechargeable battery

Battery Life:

Typically 12 hours between charges

Dimensions (mm):

260 W x 110 D x 210 H

Weight:

2 Kg

Operating Temperature: 

0 – 50°C

Cleaning:

Use moist cloth only – do not immerse in water

Delivery tube material:

PTFE and/or polyurethane recommended

Содержание GAZ-CAL

Страница 1: ...t 61 8 6108 0000 e info gastech com w gastech com Calibration Gas Generator OPERATOR S MANUAL ...

Страница 2: ...Contents Introduction 1 Specification 1 Maintenance 2 Instrument Packing List Layout 3 Operation 4 Warnings 5 Troubleshooting 5 ...

Страница 3: ...then dilutes the gas to the required concentration SPECIFICATION Concentration range Adjustable between 0 5 and 20ppm 10 or 0 25ppm whichever is greater Calibration 0 5 5ppm or 5 20ppm Flow rate 0 5 l min or 1 l min Cell Life 500ppm hours Power Supply Internal rechargeable battery Battery Life Typically 12 hours between charges Dimensions mm 260 W x 110 D x 210 H Weight 2 Kg Operating Temperature ...

Страница 4: ...ly swing front panel clear of casing 3 Lift filter from spring clip holders 4 Replace with new filter MAINTENANCE 1 Remove 4 off front panel retaining screws 2 Carefully swing front panel clear of casing 4 Replace with new filter 3 Lift filter from spring clip holders Replacing Internal Air Filter ...

Страница 5: ...ts List 1 GazCalTM 2 Shoulder Strap 3 1m of Polyurethane tubing 4 Charger 5 Operating Manual Instrument Layout 1 On Off Switch 2 Cell Life Indicator 3 Blockage LED 4 PPM Adjust Knob 5 PPM Display 6 Battery Charging Socket 7 Gas Outlet Connector ...

Страница 6: ...nt can be run from the charger 5 Ensure BLOCKAGE LED is not illuminated If the BLOCKAGE LED is illu minated check outlet pipe for blockages If the BLOCKAGE LED remains illuminated the internal air filter may need replacing The life of the air filter depends upon the atmosphere which the instrument is used in but should always be replaced every 3 MONTHS 6 Rotate PPM ADJUST knob until the desired co...

Страница 7: ...l representative or b fitting a new cell yourself new cells from Gastech come with full in structions Q Low battery symbol appears on the display A Charge instrument with GazCalTM CHARGER for eight hours Q The BLOCKAGE LED is illuminated A 1 Check that the gas outlet pipe is no longer than 1m in length as supplied by Gastech 2 Ensure that any adaptors on calibration cups fitted to the outlet do no...

Страница 8: ...EAD OFFICE 24 Baretta Road Wangara WA 6065 Phone 61 8 6108 0000 SYDNEY OFFICE 21 25 Narabang Way Belrose NSW 2085 Phone 61 2 9451 0054 BRISBANE OFFICE PO Box 349 Cannon Hill Qld 4170 Phone 61 7 3160 0901 ...

Отзывы: