2008
Gastech
Australia
Pty
Ltd
Page
34
If
the
splash
guard
is
fitted
this
will
need
to
be
removed
so
you
can
apply
the
gas
directly
to
the
sensor.
Remove
the
splash
guard
by
removing
the
two
screws
(4mm
Alan
key).
The
calibration
procedure
is
dependent
on
whether
the
sensor
is
a:
1.
Oxygen
sensor
2.
Toxic
sensor
(all
non
‐
oxygen
sensors)
Skip
to
the
appropriate
section
below
depending
on
the
sensor
installed
in
your
D
‐
Guard
TM
unit.
3.3.2.1
Calibrating
An
Oxygen
Sensor
1.
Open
the
calibration
plate
and
the
zero
calibration
menu
will
be
shown,
as
this
setting
needs
to
be
adjusted
first.
Note:
Opening
the
calibration
plate
will
lock
the
output
signal
to
the
mA
equivalent
of
fresh
air
20.9%
vol
O2.
This
will
prevent
any
alarms
being
generated
when
calibrating
the
D
‐
Guard
TM
.
2.
When
adjusting
the
zero
for
an
oxygen
sensor,
a
calibration
gas
must
be
applied.
A
calibration
plug
(Part
Number:
81
‐
1125)
will
be
required
to
apply
the
calibration
gas
directly
to
the
sensor.
Splash
Guard
Содержание D-Guard
Страница 2: ......