2008
Gastech
Australia
Pty
Ltd
Page
32
G
11
7.0
–
5.0
uA
G
12
5.0
–
3.5
uA
G
13
3.5
–
2.0
uA
G
14
2.0
–
1.5
uA
G
15
1.5
–
1.0
uA
Note:
If
the
correct
gain
setting
is
unknown,
you
can
apply
a
span
gas
to
the
sensor
and
toggle
the
gain
settings
until
you
get
a
reading
close
to
the
span
gas
concentration.
This
option
is
not
the
preferred
set
up
procedure
and
will
only
work
on
a
new
sensor
that
is
working
correctly.
Having
too
much
gain
could
result
in
excessive
sensor
drift
or
reduced
sensor
life.
17.
Press
the
“Select”
button
to
move
to
the
next
menu.
18.
The
next
screen
is
used
to
set
the
unit
of
measurement.
The
sensor
that
is
installed
determines
the
units
of
measurements.
Different
gases
are
measured
in
“%VOL”,
“ppm”
or
“%LEL”.
Use
the
“Up”
or
“Down”
button
to
toggle
between
the
various
options.
Press
the
“Select”
button
to
move
to
the
next
menu.
19.
The
next
screen
is
used
to
enable
or
disable
zero
suppression.
The
ZSUPP
function
is
part
of
the
intelligent
algorithms
of
the
D
‐
Guard
TM
.
The
D
‐
Guard
TM
advanced
software
monitors
the
sensor
performance
and
small
sensor
drifts,
which
is
a
characteristic
of
all
sensors.
This
drift
is
due
to
sudden
temperature,
humidity
and
pressure
changes
or
aging
sensors.
The
advanced
electronics
and
software
will
remove
this
insignificant
drift
and
keep
the
detector
reading
zero
or
20.9
(for
an
Oxygen
sensor).
Use
the
“Up”
or
“Down”
button
to
toggle
between
on
or
off.
Press
the
“Select”
button
to
move
to
the
next
menu.
20.
The
next
screen
will
depend
on
the
type
of
gas
the
D
‐
Guard
TM
is
setup
to
detect.
If
the
unit
is
configured
for
O
2
,
the
menu
will
show
autospan,
for
all
other
gases
the
menu
will
show
autoz.
This
Atmospheric
compensation
menu
is
one
of
the
most
advanced
features
of
the
D
‐
GuardTM.
This
logic
monitors
the
sensor
over
a
given
time
and
calculates
against
a
pre
determined
criteria
which
will
then
adjust
the
sensors
base
line
to
keep
the
sensor
in
accurate
calibration
and
eliminate
sensor
drift
which
is
a
characteristic
of
all
electrochemical
sensors
as
they
age.
This
feature
prevents
false
alarms
from
drifting
older
sensors.
Use
the
“Up”
or
“Down”
button
to
enable
or
disable
this
feature.
Press
the
“Select”
button
to
move
to
the
next
menu.