Pump Failure!
Following a pump test failure, the instrument will shut itself off.
In the event of a pump failure, perform the following steps:
1. Remove the probe assembly from the instrument and turn
the Cannonball3 back on. If the pump test passes,
proceed to step 2a below. If the pump test fails, proceed
to step 2b below.
2a If the pump test passes, then the leak is in the probe
assembly. Inspect the probe assembly and look for leaks.
Replace filters or probe part as necessary.
Figure 6.2.3 Pump/tubing assembly diagram
2b If the pump test fails, then the leak is inside the
instrument. If the leak is inside proceed to step 3.
3. Verify that the pump itself is correctly installed with the
gasket seated correctly around the base of the pump.
Retest. If the test fails proceed to step 4.
4. Remove the retention screw on the replaceable internal
filter assembly and check the filters. Replace them as
needed and reinstall the assembly. Retest. If the test fails
proceed to step 5.
5. Open up the instrument and check that all hoses are
correctly attached (see the hose diagram below) Retest.
If the pump test fails again, either replace the pump
entirely, or return the instrument to Sperian
Instrumentation.
6.2.4.
Can’t resume normal operation after a “Low Flow”
shut down
Possible causes:
1. Sample probe or internal pump filters need replacement.
2. Sample hose is kinked.
3. Sample probe and probe assembly contains fluids.
Solution(s):
Turn off Cannonball3, remove pump, disconnect sample probe
an hose assembly, allow any trapped fluids to drain; replace
filters as necessary, examine hose to make sure there are no
kinks blocking normal flow.
If these measures fail to resolve the problem, the pump itself
may be the cause. In this case, the pump assembly should be
removed from the Cannonball3 and returned to Sperian
Instrumentation.
6.3 Sample probe assembly
The sample probe handle contains moisture barrier and
particulate filters designed to remove contaminants that might
otherwise harm the instrument.
CAUTION: Never operate the Cannonball3 without the
sample probe and hose assembly. The sample probe
handle contains replaceable filters designed to block
moisture and remove particulate contaminants. If the
pump is operated without the probe assembly in place
particulate contaminants may cause damage to the pump,
sensors and internal components of the Cannonball3.
Particulate contaminants are removed by means of a cellulose
filter. The hydrophobic filter includes a 0.1
μ
m Teflon
™
barrier
which blocks the flow of moisture as well as any remaining
particulate contaminants.