27
5.6
Workflow: Set the Output of the Current Loop (optional)
To get an Output Signal representing the measured
concentration generated by the 0-20mA-Current-Loop it is
necessary to connect a suitable cable to the Signal Converter
–
I/O
Ports found at the backside of the GC-IMS-Odor Instrument as well as
to the chosen Signal Display. Please consult the GC-IMS-Odour User
Manual for more information.
1
Open the Current Loop
Settings Dialogue Box
System
>
Settings
>
Current
Loop
>
Settings…
2
Use the Value Entry Fields of
Result Value / Special
Signals -> Current
>
To set
Min
and
Max
oft he
projection of the concent-
ration range onto the current
output range (0 mA to 20 mA)
Example-Settings:
0 mg/m
3
→
4 mA
51 mg/m
3
→
20 mA