13
Anti-BUMPING/FOAMING MODE
The anti-BUMPING/FOAMING functions will decrease the set vacuum level by several torr.
Continual pressing of the Anti-BUMPING/FOAMING button will decrease the vacuum setting
an equally additional several torr, mbar or kPa.
The anti-BUMPING/FOAMING is only enabled in AUTO mode.
Press once and the vacuum decreases, press a second time and the unit goes
back to MAXIMUM VACUUM.
MENU MODE
Press MENU to access all program functions
1) Repeat
2) Program
3) Setup
4) About
5) Exit
Use the Vacuum Control knob to scroll through all the MENU options. Push the Vacuum Control
knob to select a highlighted MENU option
- 1) REPEAT
Repeat displays the previous run vacuum setting or program number. Use the Vacuum Control
knob to scroll to the Repeat option and press to repeat previous setting.
Note: The Repeat option is quickly accessible in VENT mode by pressing number “0”.
- 2) PROGRAM
Selecting the Program option allows the end user to select from one of five stored programs, enter
a new program or edit an existing one. Use the Vacuum Control knob to scroll to the Program
option and press to enter the Program menu screen.
The Program screen displays all existing stored programs in desired vacuum units and time
(minutes). Use the Vacuum Control knob to scroll up and down and press the Vacuum Control
knob to enter your selection.
If the memory location is empty the following screen will appear. Enter the values requested, first
vacuum setting (P1), first time setting in minutes (T1), second vacuum setting (P2), second time
setting in minutes (T2). If P1is entered and T1 is left blank the program will only store a vacuum
setting. If T1 and/or T2 is left blank, P2 will be set to run until unit is manually turned off.
If the values are entered correctly and have been saved, the program will redisplay the
entered values with the option to Edit, Run or Exit the program you just entered.
After each program ends with a timed value, the unit will automatically go into VENT, or
Self-Cleaning mode if Auto Shutdown is Enabled in Setup. ( see Setup)
If the memory location is full. The following screen will appear and request if you want to Edit,
Run or Exit the program. Use the Vacuum Control knob to select your option.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com