e.bloxx Communication
LOCAL-BUS PROTOCOL
HB_EBLOXX_COM_E_V12.doc
47
Gantner Instruments Test & Measurement GmbH
Get Variable All (GetVarAll)
Description:
This command is used to read the actual value of all variables of an e.bloxx equipment.
Request Telegram
L.................. Length of the data: 0x01
CHAR
CMD10 ....... Command 10 "Get Variable All": 0x0A
CHAR
FCS ............ Frame check sequence of the request
CHAR
Response Telegrams
Acknowledge:
L.................. Length of the data: 0x01 ... 0xFF
CHAR
VALUE [0]... Values 0 to be transmitted
BOOLEAN / INTEGER / LONGINTEGER /
FLOAT
...
VALUE [n]... Values n to be transmitted
BOOLEAN / INTEGER / FLOAT
n depends on the e.bloxx equipment
e.bloxx A1-x: max. n = 0x07
e.bloxx A4-x: max. n = 0x07
e.bloxx A5-1: max. n = 0x07
e.bloxx A6-2CF: max. n = 0x0F
e.bloxx A9-1: max. n = 0x07
e.bloxx D1-x: max. n = 0x07
FCS ............ Frame check sequence of the ACK response
CHAR
Not Acknowledge:
L.................. Length of the data: 0x01
CHAR
ErrorCode... Command not available/invalid: 0x01
CHAR
FCS ............ Frame check sequence of the NAK response
CHAR
SYNC2 ADDR
L
CMD10 FCS
ACK
ADDR
L
VALUE [0] VALUE [1] ...
VALUE [n] FCS
NAK
ADDR
L
ErrorCode FCS