Operational Method

 

3.1

 

Turn  on  the  power.  Open  the  switch  by  pressing

.  The  displaying  windows 

displays  from  “F----0”to  “F----9”  and  after  a  certain  time  for  stabilizing, 

0

·

0

” appears; Then preheats for 15 minutes. Instability of the displaying at 

every beginning is normal; after a certain time, the displaying will stabilize. 

 
3.2

 

With empty scale platform, the displaying deviates from zero, pressing “TARE” 
key, then the displaying will return to zero. By Taring first place the container 
on the platform, after stabilizing, press “TARE”;the scale displays ”zero”. Then 
place the article in container, at this time, it shows the net weight of the article. 
Remove  the  article  and  the  container.  The  scale  shows  a  negative  value  of 
container weight. Still press “TARE” to return to “zero” 

 

 
3.3

 

The units can be changed between “kg” and “lb”. Press 

  to change the units 

during weighting 

 

3.4

 

  To use the balance in dark place, press the 

 

key to active the background 

light from display. 

 

3.5

 

  To print the weighing data with extern printer, just press “PRINT” key. 

 
3.6

 

  If the scale has not been used for a long time or just bought, the scale should 
be  calibrated.  Firstly,  the  scale  is  fully  preheated  (over  15  min.)  with  empty 
scale  platform.  Press  “TARE”  key  and  then  press

,  the  displaying  window 

shows  “CXXX”  and the balance  will  automatic  calibrate (XXX should be the 
weight of standard weights). At  the same  time,  only place calibrating weights 
on  the  scale  platform.  After  stabilizing,  the  scale  displays  the  weight  of  the 
calibrating weights; till now the calibration is completely finished and the scale 
can  be  normally  used.    If  the  Display  shows  the  message  “C----F”  by  press 
“CAL”, but display, which indicates the Zero point is not stable. Then re-press 
“TARE”  to  make  “zero”  point  stable;  and  re-press  the  “CAL”  to  carry  out 
calibration. 

 

Содержание TC50K

Страница 1: ...1 41564 Kaarst Germany Operation Manual For Models TC50K TC60K TC100K TC150K TC200K TC300K TC60KA TC75KA TC100KA TC150KA TC300KA To guarantee the best performance of the device please read this operat...

Страница 2: ...Specification and technical data 3 Operation method 4 Attachment Configuration and description of function Change 5 Data transmission 6 Structure 7 Caution for Operation 8 Trouble shooting 9 Packing...

Страница 3: ...apting to the adverse environment and long hours continuous work With double power supply AC DC With auto calibration With Count function the counting of Object with large quantity simplified Unit con...

Страница 4: ...kg 0 150 kg 0 200 kg 0 300 kg Plattform Dimension 520 x 410 mm Volume 970 B x 515 T x 205 H mm Power Supply AC 230V 50Hz 60Hz 4W Rechargeable Battery DC 6V 4Ah 20HR Use Temperature 0 40 Use Humidity 8...

Страница 5: ...till press TARE to return to zero 3 3 The units can be changed between kg and lb Press to change the units during weighting 3 4 To use the balance in dark place press the key to active the background...

Страница 6: ...to change the configuration and function the display shows C1 and the relevant number which can be changed Pressing can change the parameters pressing TARE can change the values of parameters 4 1 C1 s...

Страница 7: ...enter 2unit 7units 3unit 1 unit 1 unit 3 8 4 Check bits no check bits 3 8 5 Output pins content 9 core socket 2 pins RXD 3 pins TXD 5pins GND 3 8 6 Computer can control balance though RS232 The order...

Страница 8: ...8 P a g e 6 Structure 1 TARE key 2 Background light key 3 COUNT key 4 Kg lb key for unit change 5 PRINT key 6 ON OFF key 7 Special signal interface 8 RS 232 interface 9 Power supply...

Страница 9: ...use of electronic balance without power supply 7 4 Please use separate and individual plug for the balance to avoid any possible interference or disturbance 7 5 Place the scale on a firm table for op...

Страница 10: ...e message F H 1 Object weight beyond capacity 3 Error by sensor 1 remove the object 2 restore the fabric configuration The message C L by Calibration Error by sensor restore the fabric configuration T...

Страница 11: ...ing the collection systems available If electrical appliances are disposed of in landfills or dumps hazardous substances can leak into the groundwater and get into the food chain damaging your health...

Отзывы: