P.S. Only use the provided adapter or the same specifications of the adapter,
otherwise there will be dangerous.
3.
Operational Method
3.1
Turn on the power. Open the switch by pressing
. The displaying windows
displays from “F----0”to “F----9” and after a certain time for stabilizing,
“
0
·
0
” appears; Then preheats for 15 minutes. Instability of the displaying at
every beginning is normal; after a certain time, the displaying will stabilize.
3.2
With empty scale platform, the displaying deviates from zero, pressing “TARE”
key, then the displaying will return to zero. By Taring first place the container
on the platform, after stabilizing, press “TARE”;the scale displays ”zero”. Then
place the article in container, at this time, it shows the net weight of the article.
Remove the article and the container. The scale shows a negative value of
container weight. Still press “TARE” to return to “zero”
3.3
The units can be changed between “kg” and “lb”. Press
to change the units
during weighting
3.4
To use the balance in dark place, press the
key to active the background
light from display.
3.5
To print the weighing data with extern printer, just press “PRINT” key.
3.6
If the scale has not been used for a long time or just bought, the scale should
be calibrated. Firstly, the scale is fully preheated (over 15 min.) with empty
scale platform. Press “TARE” key and then press
, the displaying window
shows “CXXX” and the balance will automatic calibrate (XXX should be the
weight of standard weights). At the same time, only place calibrating weights
on the scale platform. After stabilizing, the scale displays the weight of the
calibrating weights; till now the calibration is completely finished and the scale
can be normally used. If the Display shows the message “C----F” by press
“CAL”, but display, which indicates the Zero point is not stable. Then re-press
“TARE” to make “zero” point stable; and re-press the “CAL” to carry out
calibration.