3. Operational Method

        3.1 Turn on the power.  Open  the  switch.  The  displaying  windows 
displays from “F----1” to “F----9” and after a certain time for stabilizing, 
“0.0” appears; Then preheats for 15 minutes. Instability of the displaying 
at every beginning is normal; after a certain time, the displaying will stabilize.
        3.2 With  empty  scale  platform,  the  displaying  deviates from zero, 
pressing “TARE” key, then the displaying will return to zero.
        3.3 If the scale has not been used for along time or just bought, the 
scale should be calibrated. Firstly, the scale is fully preheated (over 15 min.) 
with empty scale platform. Then press “CAL”, the displaying windows hows 
“C-XX--” and the balance will automatic calibrate (XX should be the weight 
of standard weights. For instance, “C-10-” means to place the 10kg test 
weights). At the same time, only place calibrating weights on the scale 
platform. After stabilizing, the scale displays the weight of the calibrating 
weights. The weight  unit  is  “kg”.  The  indicator  lightens;  till  now  the 
calibration is completely finished and the scale can be normally used. If 
press “CAL”, but display “C----F”, which indicates the Zero point is not 
stable. Then re-press “TARE” to make “zero” point stable; and re-press 
the “CAL” to carry out calibration.
        3.4 If weight of a weighed article exceeding the scale measuring range, 
the scale will warn “F----H”.
        3.5 If needs to tare,  first place  the  container  on  the platform,  after 
stabilizing, press “TARE”; the scale displays ”zero”. Then place the article 
in container, at this time, it shows the net weight of the article. Remove the 
article and the container. The scale shows a negative value of container 
weight. Still press “TARE” to return to “zero”

3.6 Count

        3.6.1 Choose samples. To precisely count the articles, firstly the quantity 

of samples for counting should be chosen in according with the weights of 
the articles. The quantity of samples can be chosen from “1-10-20-50-100”. 
For a small number of articles, the larger quantity of samples, the more 
precise.
        3.6.2 Place the samples on the pan, the balance will show the total 
weights of samples. Then press “Count”, balance will show “1”, and unit 
will show “pcs”, which indicate that balance is now counting. And all the 
samples are one unit. Then press “Count”, the displaying will show between
1-10-20-50-100. The quantity of articles must be the same with the quantity 
of samples, and place the similar articles. The total amount of objects is 
showed. Pressthe “Count” again back to the normal situation.
        3.7 Unit Change: Press “kg/lb”, the units will be changed between 
“kg” and “lb”. The indicator willlighten.
        3.8 Data Output 
        Balance is targeting data output interface RS232C and can be directly 
connected to the printer. The following soft wares are needed, when connects 
to computer:
        3.8.1 Transmission format: serial asynchronous
        3.8.2 Serial communication protocol:
                 Baudrat: 600,
                 data: 8 units,
                 starting: 1 unit, 
                 ending:1 unit
        3.8.3 Data format: 
                 starting         data        symbol        ending        enter
                  2 unit         7 units       3 unit          1 unit         1 unit
        3.8.4 Calibration: no calibration
        3.8.5 Output pins content (9-core socket): 2 pins: RXD   3 pins: TXD
5 pins: GND

Содержание TC3K

Страница 1: ...K TC6K TC10K TC20K TC30K TC10KB TC3KA TC6KA TC10KA TC20KA TC30KA Operation Manual G G GmbH Gielenstr 65 69 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specifications and Performance indexes 3 Operational Method 4 Structure 5 RS232 Data output demonstration 6 Cautions for operation 7 The Warranty 8 Trash 9 Contact ...

Страница 2: ...e can even remote controlled by computer 1 8 With counting function In general with high precision high stability high reliability and high efficiency the G G electronic balances has widely replace the traditional mechanical tray balances The G G electronic balances have been widely used in all callings in different kinds of measurement and test Those have played an important role in improving pro...

Страница 3: ...le displays zero Then place the article in container at this time it shows the net weight of the article Remove the article and the container The scale shows a negative value of container weight Still press TARE to return to zero 3 6 Count 3 6 1 Choose samples To precisely count the articles firstly the quantity of samples for counting should be chosen in according with the weights of the articles...

Страница 4: ...nner not specified by this instruction the projection provided by the equipment may be impaired 7 The Warranty 7 1 We provide warranty within the scope of the statutory regulations however we are entitled to either repair or replace the defectiv eproduct 7 2 Within two years after purchase the producer offers a free repair service to all G G Product which become defect under proper usage without h...

Страница 5: ... telephone number of the buyer so that we can mail the repaired product back in time 8 Trash Meaning of crossed out wheeled dustbin Do not dispose of electrical appliances as unsorted municipal waste use separate collection facilities Contact you local government for information regarding the collection systems available If electrical appliances are disposed of in landfills or dumps hazardous subs...

Отзывы: