background image

3

4

“G&G”

        3.4 The units can be changed between “g”, “ct”, “ozt”, “oz”, “GN”, 
“dwt”, “lb”. Press “Units” to change the units during weighting
        3.5 If needs to tare,  first place the  container  on  the  platform,  after 
stabilizing, press “TARE”; the scale displays “zero”. Then place the article 
in container, at this time, it shows the net weight of the article. Remove the 
article and the container. The scale shows a negative value of container 
weight. Still press “TARE” to return to “zero”
        3.6 .Count function
        3.6.1 Change the balance modus into count modus according to the 
Attachment. 
        3.6.2 Choose samples. To precisely count the articles, firstly the quantity 
of samples for counting should be chosen in according with the weights of 
the articles. The quantity of samples can be chosen from “1-10-20-50-100”. 
For a small number of articles, the larger quantity of samples, the more 
precise. 
        3.6.3 Place the samples on the pan, the balance will show the total 
weights of samples. Then press “Count”, balance will show “1”, and unit 
will show “pcs”, which indicate that balance is now counting. And all the 
samples are one unit. Then press “Count”, the displaying will show between 
1-10-20-50-100. The quantity of articles must be the same with the quantity 
of samples, and place the similar articles. The total amount of objects is 
showed. Press the “Count” again back to the normal situation.

4. Attachment: settings and meanings

        Turn on the Scale by pressing “CAL” key, in order to change the con-
figuration and function. the display shows C1 and the relevant number, 
which can be changed. Pressing “CAL” can change the parameters; pressing 
“TARE” can change the values of parameters.
        4.1  C1------ sensitivity setting 0 1 2 3 4, the larger number is, the less 
sensible but the better stable. Default setting is 2 or 1
        4.2  C2------filtering strength setting 0 1 2 3, the larger number is, the 

lower reaction speed, but the better stable. Default setting is 2 or 1
        4.3  C3------Baud rate setting 2(600) 3(1200) 4(2400) 5(4800) 6(9600). 
Default  setting is 2
        4.4  C4------communication setting. “TARE” changes the lower unit. 
“COUNT” changes the higher units. This is the first data during commun-
ication. Default setting is 27.
        4.5  C5------units change. 0 is for units changing and 1 for printing. 
Default setting is 0. Counting is 2.
        4.6  Re-pressing “CAL” after setting. Then the balance is auto calibr-
ated to 0.
        4.7  Press “TARE” and do not loose till “F----3” (Balance should be 
switched on).
        All the parameters can be reset to default settings. Re-calibration before 
using.

5. Data Transmission

        Balance is targeting data output interface RS232C and can be directly 
connected to the printer. The following soft wares are needed, when connects 
to computer:
        3.8.1 Transmission format: serial asynchronous
        3.8.2 Serial communication protocol: Baud rat: 600, data: 8 units, 
starting: 1 unit, ending: 1unit
        3.8.3 Data format: 
                 Starting spaces    data     symbol    ending      enter
                          2unit         7units      3unit      1 unit      1 unit
        3.8.4 Check bits: no check bits
        3.8.5 Output pins content (9-core socket): 2 pins: RXD   3 pins: TXD   
5pins: GND
        3.8.6 Computer can control balance though RS232. The orders to 
balance are:
           1BH+70H: sending data

Содержание JJ100B

Страница 1: ... JJ2000B Operation Manual G G GmbH Koenigstr 34 41460 Neuss Germany G G GmbH Content 1 Overview 2 Specification and technical data 3 Operation method 4 Attachment Configuration and description of function Change 5 Data transmission 6 Structure 7 Caution for Operation 8 Trouble shooting 9 Packing list 10 Disposal 11 Warranty 12 Contact ...

Страница 2: ...es Depending on the difference of range and resolution G G Brand JJ B series electronic scales are classified to 4 types as shown Model JJ100B JJ200B JJ1500B JJ2000B Capacity 110g 200g 1500g 2000g Resolution Ratio 0 001g 0 001g 0 01g 0 01g Calibration Graduation 10d 10d 10d 10d Tare Scope 2000g 3000g Calibrating Weight 2000g 2000g Dimension of Platform Ф135mm Volume 325 W X 211 D X 111 H mm Power ...

Страница 3: ...the con figuration and function the display shows C1 and the relevant number which can be changed Pressing CAL can change the parameters pressing TARE can change the values of parameters 4 1 C1 sensitivity setting 0 1 2 3 4 the larger number is the less sensible but the better stable Default setting is 2 or 1 4 2 C2 filtering strength setting 0 1 2 3 the larger number is the lower reaction speed b...

Страница 4: ...ectronic balance without power supply 7 3 Place the scale on a firm table for operation and make sure that the operating environment of the balance is free of great vibration and interfe rence of power supply and of corrosive gases and liquids 7 4 Ensure the warm up time after power on 7 5 Warning If the equipment is used in a manner not specified by this instruction the perfection provided by the...

Страница 5: ... 41460 Neuss Germany Tel 49 2131 609 7271 Fax 49 2131 609 7273 Email info gandg de Website www gandg de Remarks Option by JJ100B JJ200B Quantity 1 1 1 1 1 Article Description Operation manual Scale pan Power cable Calibration weight Wind shield restore the fabric configuration 1 remove the object 2 restore the fabric configuration restore the fabric configuration Press the TARE key with empty pan ...

Отзывы: