Chapter 3: Unpacking and Installation
–
Complete Unit--Power-up test
3 - 4
The external power supply provided with the Stir/Purge is rated for operation from 100 to 240 V AC, at
frequencies from 47 to 63 Hz. It should therefore be useable worldwide.
The Stir/Purge external supply is supplied with a line cord suitable for use in the United States. In other
countries, you may have to replace the line cord with one suitable for your electrical outlet type. You must
always use a line cord with a CEE 22 Standard V female connector on the apparatus end of the cable. This is
the same connector used on the US standard line cord supplied with your Stir/Purge. See Chapter 1 for specific
safety information regarding line cord selection.
The DC output cord from the external power supply plugs into the POWER IN jack on the rear panel of the
Stir/Purge (see below):
Power-up test
Before you make any other connections to your Stir/Purge, power up the Stir/Purge and watch the blue POWER
LED indicator on the front panel of the Stir/Purge:
After connecting DC power to the Stir/Purge, turn on the POWER switch on the front panel of the Stir/Purge
(see Figure 5 - 1).
Watch the POWER LED as the Stir/Purge starts up. This LED should remain illuminated. The status of the other
LED indicators is not important at this time.
Caution: If the POWER LED goes on, then turns off and stays off, the
Stir/Purge is not working properly! Contact Gamry Instruments or your local Gamry
Instruments representative as soon as possible if this power-up test fails.
To AC wall outlet (mains)
To POWER IN
connector on
back of Stir-
Purge
Insert line
cord into
power supply
12 V DC input power
from external supply
Содержание VistaShield Stir/Purge
Страница 1: ...VistaShield Stir Purge Operator s Manual...
Страница 2: ......
Страница 6: ...iv...
Страница 14: ...Chapter 1 Safety Considerations RoHS Compliance 1 6...
Страница 22: ...Chapter 3 Unpacking and Installation Complete Unit Gas fitting installation 3 6...
Страница 26: ...Disassembly of the Stir Purge from the VistaShield 4 4...
Страница 36: ...Certificate of Conformance 6 4 Certificate of Conformance...
Страница 38: ...Appendix C Specifications 6 6...
Страница 40: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com...