5.  Entering and Checking the New Tuning Parameters 

a)

 

The values required by the TDC4 for tuning are PRoP, RESt, and RAtE. 

b)

 

Enter the values calculated above into the TDC4 by accessing the CNFG, OUt 1, CtRL menu.   

c)

 

Set the controller back to Automatic Mode by changing SELF mode back to dSbL. 

d)

 

Enter a setpoint that is 10 to 20 

°

C above the current temperature. 

e)

 

Rerun the "TDCTUNE.EXP" script.  A smooth rise in temperature should be seen.  A overshoot of 
1 or 2 

°

C is acceptable. 

 
 

 

Содержание TDC4

Страница 1: ...TDC4 Temperature Controller Operator s Manual Copyright 2010 2018 Gamry Instruments Inc All rights reserved Printed in the USA Revision 5 41 July 1 2015...

Страница 2: ...me without incurring any obligation to install same on TDC4s previously purchased All system specifications are subject to change without notice There are no warranties which extend beyond the descrip...

Страница 3: ...us about a problem provide as much information as possible If you are having problems in installation or use of this TDC4 Temperature Controller it would be helpful if you called from a phone next to...

Страница 4: ...ework Software Originally CMS100 Software Copyright 1989 2015 Gamry Instruments Inc CPT110 Software Originally CMS110 Software Copyright 1992 2015 Gamry Instruments Inc Explain Computer Language Copyr...

Страница 5: ...2 AC Line Connection 2 2 Power Up Check 2 3 RS232 Cable 2 3 USB to COM Converters 2 4 Connecting the TDC4 to a Heater and or Cooler 2 4 Connecting the TDC4 to an RTD Probe 2 4 Cell Cables from the Po...

Страница 6: ...ive grounding can be rendered ineffective if the TDC4 is damaged in shipment Do not operate damaged apparatus until its safety has been verified by a qualified service technician Tag a damaged TDC4 to...

Страница 7: ...or both US line cords and European line cords that follow the harmonized wiring convention Table 1 1 Line Cord Polarities and Colors If you have any doubts about the line cord for use with your TDC4 p...

Страница 8: ...it is first powered up Warning The switched electrical outlets on the TDC4 rear panel must always be treated as on whenever the TDC4 is powered Remove the TDC4 line cord if you must work with a wire...

Страница 9: ...are now generally available from computer supply houses and electronics tool suppliers An antistatic floor mat may also help particularly if a carpet is involved in generating the static electricity...

Страница 10: ...carrier Warning The protective grounding can be rendered ineffective if the TDC4 is damaged in shipment Do not operate damaged apparatus until its safety has been verified by a qualified service tech...

Страница 11: ...devices one to the 3 pin Gamry connector and one to the terminal strip Unplug the RTD from its connector if you connect any sensor to the input terminal strip You must reconfigure the controller for...

Страница 12: ...obe installed the upper display should show a line containing the characters OPN Once the unit has powered up correctly turn it off prior to making the remaining system connections RS232 Cable The TDC...

Страница 13: ...om Output 1 on the rear panel of the TDC4 This output is a CEE 22 Standard V female connector Electrical cords with the corresponding male connector are available worldwide Please check that the fuses...

Страница 14: ...hich is configured by means of user entered parameters Note Please refer to the Omega Documentation supplied with your TDC4 for information about the various controller parameters Do not change a para...

Страница 15: ...2 6...

Страница 16: ...at has a Gamry potentiostat installed in the system and currently connected If you do not have a potentiostat in the system the script will show an error message and terminate before it outputs anythi...

Страница 17: ...ating mantle is moderately slow An internal heater such as a cartridge heater is often quicker Tuning the TDC4 Temperature Controller Overview Closed loop control systems such as the TDC4 must be tune...

Страница 18: ...using close to one hundred settings fully listed in Appendix B which can be entered by a complicated user interface using the four buttons on the front of the TDC4 We anticipate that anyone using this...

Страница 19: ...many electrochemical cells is too slow for auto tuning You cannot auto tune if a 5 C increase or decease in the system temperature takes more than ten minutes In most cases auto tune on an electrochem...

Страница 20: ...d using the same cell that will be used for real experiments You should try to duplicate your experimental conditions as closely as possible 3 The script that starts the auto tune process is called TD...

Страница 21: ...ware and verify CPT110 operation prior to using this script 1 Setup the System a Set up the cell and its stirrer The cell must be filled with liquid with roughly the same thermal properties and volume...

Страница 22: ...b Restart the TDC4 Manual Tune exp Framework script c Record a few hundred seconds of base line data d Enter a duty cycle that is about 1 5 times higher than current duty cycle For example if the prev...

Страница 23: ...File Open from the Excel menu In the resulting dialog box select the TUNE DTA file in your Gamry Framework data directory Select Ok NOTE The data curve resulting from the TDCTUNE experiment is normall...

Страница 24: ...ion in determination of the deadtime see below If you select too few points the slope can be difficult to determine f Select Insert Chart On This Sheet from the Excel menu g Drag the cursor over a bla...

Страница 25: ...of Tuning Parameters a Using the chart from step 3 estimate the dead time L This is the time prior to the start of the linear increase in the temperature If a line approximating the linear increase i...

Страница 26: ...ata in Figure 3 7 this becomes R 7 7 x 10 4 C sec 4 1 9 x 10 4 C sec output d Calculate the proportional gain Kc from Kc 1 RL in our example Kc 1 1 9 x 10 4 300 17 Output C e The TDC4 s proportional b...

Страница 27: ...values calculated above into the TDC4 by accessing the CNFG OUt 1 CtRL menu c Set the controller back to Automatic Mode by changing SELF mode back to dSbL d Enter a setpoint that is 10 to 20 C above t...

Страница 28: ...ripheral TDC4 to computer 3 TX Input Data from computer to peripheral 5 Common N A Ground connection A cable suitable for connection of a TDC4 to a standard PC COM port was included with your TDC4 shi...

Страница 29: ...0 1 0 1 0 0 10 0 20 NA RW 07 RdG dEC FFFF FFF F FF FF F FFF FFF F RW 08 tEMP F C C RW 08 FLtR 0001 0002 0004 0008 0016 0032 0128 0004 RW 08 IN Rd if Process selected IN 1 W V NA Rd 1 W V NA IN 2 W V N...

Страница 30: ...St W V 685 0 RW 18 RAtE W V 109 0 RW 19 CyCL W V 01 RW 1A dPNG 0000 0001 0002 0003 0004 0005 0006 0007 0003 RW 0D dEAd if ON OFF selected W V 014 0 RW 17 OUt2 CtRL ON OF PId ON OF RW 0D ActN dRct RVRS...

Страница 31: ...W 1F _LF_ NO YES YES RW 1F ECHO NO YES NO RW 1F StNd 232C 485 232C RW 1F ModE CMd CoNt CMd RW 1F SEPR SPCE _cR_ SPCE RW 1F dAt F StAt NO YES YES RW 20 RdNG NO YES YES RW 20 PEAK NO YES YES RW 20 VALY...

Страница 32: ...D 2 5 Omega CSi8D Temperature Controller 2 1 3 1 Output 1 2 4 Output 2 2 4 Parameters Operating 3 2 power line transient 1 4 Product Safety 1 1 protective grounding 1 1 RFI 1 3 RS232 cable 2 3 RTD 2 4...

Отзывы: