Gamry Instruments Reference 30k Booster Скачать руководство пользователя страница 28

Installation – Power-Up Test 

28 

The correct connections are shown in Figure 3-1. 

Figure 3-1 

Connections to the Cal Side of the Test & Cal Cell 

 

Power-Up Test 

The power switches on the Reference 3000 potentiostat and the Reference 30k Booster can be turned on in 
any order. 

When power is applied, the blue 

Power

 LED indicators on both the Reference 3000 and the Reference 30k 

should illuminate. If your computer is powered and connected to the Reference 3000’s USB, the 

USB

 LED on 

the Reference 3000 should glow green. 

No other LEDs should be illuminated. If either instrument has its red 

Overload

 LED lighted, turn off the power 

and recheck all the connections. An open fuse in the 

Cal

 Cell could cause an 

Overload

 condition. 

Calibration 

Run the Gamry Framework software. In an earlier installation step, you already had the Reference 3000 
running, so it should already be installed as a Gamry Device. If you get an error message when the Framework 
software runs, contact Gamry Technical Support.   

Assuming that the Framework software is running, select 

Experiment > Utilities > Calibration

 on the 

Framework menu. A dialog box like the one shown in Figure 3-2 should appear: 

Содержание Reference 30k Booster

Страница 1: ...Reference 30k Booster Operator s Manual Copyright 1991 2021 Gamry Instruments Inc Revision 1 7 August 24 2021 988 00021...

Страница 2: ......

Страница 3: ...cant engineering time on our part can be performed on a contract basis Contact us with your requirements Limited Warranty Gamry Instruments Inc warrants to the original user of this product that it sh...

Страница 4: ...at might appear Copyrights Reference 30k Booster Operator s Manual copyright 1991 2021 Gamry Instruments Inc all rights reserved Gamry Framework copyright 1989 2021 Gamry Instruments Inc all rights re...

Страница 5: ...e 16 Chapter 2 Introduction 17 Compatibility with Older Reference 3000 Potentiostats 17 Cable Connections from Reference 3000 to the Reference 30k Booster 18 Analog Cable 18 Digital Cable 18 Theory of...

Страница 6: ...al Clips 41 General Suggestions for High current Connections 41 Large diameter Threaded Connections 42 Appendix A Reference 30k Booster Specifications 43 Control Amplifier 43 30 A Current Range 43 Sys...

Страница 7: ...paratus is only safe from electrical shock hazards if the case of the apparatus is connected to a protective earth ground In the Reference 30k Booster this protective ground connection is made via the...

Страница 8: ...minal Line voltage Setting Nominal Setting Allowed Range Fuse Rating Schurter P N 115 90 to 130 VAC 12 5 A 250 VAC 0034 3128 240 215 to 264 VAC 6 3 A 250 VAC 0034 3125 The instrument automatically swi...

Страница 9: ...nged with the AC line mains cord plugged into the Reference 30k Booster 2 Locate the AC input module on the lower left hand corner of the Reference 30k Booster rear panel It can be identified by the p...

Страница 10: ...ambient temperatures between 0 C and 40 C Be careful when operating the Reference 30k Booster in an enclosed space such as an enclosed relay rack or NEMA enclosure Caution The temperature within the e...

Страница 11: ...ing Shipping Same as storage plus Acceleration Maximum 30 G Operation Ambient Temperature 10 C to 40 C Relative Humidity Maximum 90 non condensing Spill Hazard Cooling of the Reference 30k Booster req...

Страница 12: ...nce 30k Booster Placing your Reference 30k Booster on a static control work surface may help Static control work surfaces are now generally available from computer supply houses and electronics tool s...

Страница 13: ...The fuses protect the cell from excess current from two possible error conditions 1 improper grounding 2 electronics failure in the booster The grounding conditions that create the problem are discuss...

Страница 14: ...system They are present to prevent damage to the Reference 3000 in two possible error conditions The first error condition requires a booster malfunction The malfunction must cause one of the two swit...

Страница 15: ...ont panel of the Reference 30k Booster glows when the EMERGENCY SHUTDOWN switch is in the tripped position The system will not run experiments until the switch is reset and the indicator light is off...

Страница 16: ...ergy A short circuit across the cell terminals can cause large currents and hazardous conditions The dangers can include fire explosion and emission of hazardous chemicals and can be severe enough to...

Страница 17: ...rement of impedances from 10 to 10 k without changing system connections Applications for a system containing a Reference 30k Booster include research quality control and production in these areas Pre...

Страница 18: ...ence 3000 s current measurement circuit Of course adding an additional measurement range is useless if the system cannot control cell voltage or current The Reference 30k Booster also contains a power...

Страница 19: ...ier comes from the voltage on the Reference 3000 s Counter electrode lead A second switch connects the working electrode to the 30 A current measurement circuit The output of this current measurement...

Страница 20: ...Introduction Theory of Operation 20 Figure 2 1 Simplified Range switching Reference 30k Booster...

Страница 21: ...Reference 3000 s current ranges or can be measured on the Reference 30k Booster s 30A range In either case the current flows between the high current terminals on the front panel of the Reference 30k...

Страница 22: ...3 m 6 W resistor and a Cal side that is a 200 m 1 W resistor The nominal tolerance of the resistors is 1 The Test side of the fixture can handle the full 30 A output of the Reference 30k Booster While...

Страница 23: ...Cal cell 3 m and 200 m 5 985 00116 Cable Ref3000 to Ref30k Working Counter 15 pin D Male to Female 6 985 00120 Cable Ref3000 to Ref30k Expansion Interface 25 pin D Male to Female 7 630 00023 Fuse Blad...

Страница 24: ...sitting on a flat surface Operation in other positions can result in restricted airflow and overheating of the circuitry We do not recommend placing solid objects near the Reference 30k The Reference...

Страница 25: ...witches to use any power line mains voltage in these ranges however safe full power operation in each range requires different AC power line fuses as noted in Table 1 2 Connections between Reference 3...

Страница 26: ...on green terminal and red wires on red terminal Cal Cell Connections Connect the other end of the Working Counter cable to the Cal Calibration side of a Reference 30k Booster Test Cal cell This Test C...

Страница 27: ...d red wires on the terminal labeled Red Make sure your connections are to the Cal side of the fixture The other three connections to the test cell come from the Reference 3000 Sense Cable They are the...

Страница 28: ...Reference 3000 should glow green No other LEDs should be illuminated If either instrument has its red Overload LED lighted turn off the power and recheck all the connections An open fuse in the Cal C...

Страница 29: ...n the Calibration Cell causes calibration failures Optional 30 A Test If you wish to further test your system you can switch the cell cabling to the Test side of the Test Cal cell This side is a 3 m 6...

Страница 30: ...cm2 and an area of 1 0 cm2 After you click the OK button a runner window displays a current and voltage versus time curve that looks similar to Figure 3 4 Figure 3 4 Current versus Time Curve for 30...

Страница 31: ...s fuses and connectors are discussed in detail in this chapter Figure 4 1 Front Panel of the Reference 30k Booster Front panel Indicators There are six indicator LEDs on the Reference 30k Booster s fr...

Страница 32: ...goes on and stays on this can indicate a dangerous condition especially when the cell cannot handle high power Over Temp LED This is a red LED which lights when the Booster heat sink exceeds a tempera...

Страница 33: ...EMERGENCY SHUTDOWN The emergency shutdown control is a big red button on the right side of the Reference 30k Booster s front panel It gives you an easy method to turn off an experiment EMERGENCY SHUTD...

Страница 34: ...8 16 bolt It is normally connected to the cell with a heavy gauge low inductance cable The standard length cable is 90 cm long Its Gamry P N is 985 00122 A Reference 3000 Reference 30k Booster system...

Страница 35: ...ector The Reference 3000 Counter Working connector on the front panel of the Reference 30k Booster connects two analog signals between the booster and the Reference 3000 The signals are the Reference...

Страница 36: ...ck and the power switch The fuses and fuse selection are described in Chapter 1 You must use fuses appropriate for the AC line voltage mains in your area Line cord selection is also described in Chapt...

Страница 37: ...mon voltage reference point for the Reference 30k Booster s circuitry It floats with respect to the Protective Ground A strap is provided to connect the Protective Ground and System Ground Tests on el...

Страница 38: ...Reference 3000 Reference 30k Booster system make this connection with a 25 pin D to 25 pin D cable Gamry P N 985 00120 Caution The Gamry 985 00120 cable contains special shielding and grounding Do no...

Страница 39: ...ge and conversion devices Large high rate batteries often have cell impedance less than 1 m Test set ups for these devices require contact and lead resistances less than 10 Fortunately you can often e...

Страница 40: ...1 shows a half of a reasonably good 4 terminal connection with two alligator clips to a battery terminal Each clip contacts the terminal independently and the sense clip is closest to the battery The...

Страница 41: ...eries and supercapacitors are generally made using large diameter bolts like those found on the Reference 30k front panel This can lead to very low resistance contacts Remember that copper is an excel...

Страница 42: ...ake contact with a larger diameter terminal you can Remove the ring terminal of the 985 00122 cable and replace it with a larger diameter terminal This may require the services of a professional elect...

Страница 43: ...static ZRA Allowed Current Ranges 30 3 0 3 0 03 A Mechanical Dimensions w x h x d 37 x 23 x 44 14 5 x 9 x 17 5 cm inch Weight 16 35 kg lbs Power Input Voltage 90 to 263 VAC Power max 1200 W A potentio...

Страница 44: ......

Страница 45: ...rotective Ground is available on a binding post on the rear of the Reference 30k Booster This binding post is green and identified with the standard symbol for protective earth ground The internal cir...

Страница 46: ...a contributor to ground issues System Ground The Expansion Interface cable between the Reference 3000 and Reference 30k Booster connects the System Grounds of the two instruments All control signals p...

Страница 47: ...d contact Gamry Instruments technical support staff Figure 5 3 shows the ground jumper in place Figure 5 3 Jumper Connection between Protective Ground and System Ground When earth grounding the System...

Страница 48: ...lated from Earth Ground Most laboratory electrochemical cells are isolated from earth ground A glass cell on a bench top is an obvious example Commercial packaged batteries and supercapacitors are usu...

Страница 49: ...d and the cell voltage In all cases the voltage will be less than 24 volts Limits on Ground Voltage The Reference 3000 contains TVS surge suppressors that limit the voltage difference between the Refe...

Страница 50: ......

Страница 51: ...owing standards as defined in report number R0540 000 from Ergonomics Inc for safety analysis and report number R 1571P from Radiation Sciences Inc for EMC test and analysis EMC Standards Title Class...

Страница 52: ...Certifications 52 Certificate of Conformance...

Страница 53: ...is based upon compliance with the following directives Electromagnetic Compatibility Regulations 2016 Electrical Equipment Safety Regulations 2016 EMC Regulations Title Class Criteria EN 61326 1 2013...

Страница 54: ......

Страница 55: ...driver 26 27 operation 11 operator errors 16 Overload 32 Overload LED 28 32 Over Temp 32 Over Temp LED 32 pass through mode 14 17 19 21 22 35 37 plastic barrier 16 power input connector 36 power LED...

Отзывы: