Installation – Power-Up Test
28
The correct connections are shown in Figure 3-1.
Figure 3-1
Connections to the Cal Side of the Test & Cal Cell
Power-Up Test
The power switches on the Reference 3000 potentiostat and the Reference 30k Booster can be turned on in
any order.
When power is applied, the blue
Power
LED indicators on both the Reference 3000 and the Reference 30k
should illuminate. If your computer is powered and connected to the Reference 3000’s USB, the
USB
LED on
the Reference 3000 should glow green.
No other LEDs should be illuminated. If either instrument has its red
Overload
LED lighted, turn off the power
and recheck all the connections. An open fuse in the
Cal
Cell could cause an
Overload
condition.
Calibration
Run the Gamry Framework software. In an earlier installation step, you already had the Reference 3000
running, so it should already be installed as a Gamry Device. If you get an error message when the Framework
software runs, contact Gamry Technical Support.
Assuming that the Framework software is running, select
Experiment > Utilities > Calibration
on the
Framework menu. A dialog box like the one shown in Figure 3-2 should appear: