Safety Considerations – EMERGENCY SHUTDOWN Switch
15
EMERGENCY SHUTDOWN Switch
Another safety feature built into the Reference 30k Booster is the
EMERGENCY SHUTDOWN
switch. This large
red knob on the front panel of the Reference 30k Booster allows the user to immediately shut off all electrical
power flowing into or out of the cell from the Reference 30k Booster. Pressing on this obvious switch causes
the switch to open. It remains open until reset by the user.
A red LED indicator on the front panel of the Reference 30k Booster glows when the
EMERGENCY
SHUTDOWN
switch is in the tripped position. The system will not run experiments until the switch is reset and
the indicator light is off.
The
EMERGENCY SHUTDOWN
switch is normally activated if you see a serious problem in your test cell, such
as boiling of the cell electrolyte, gas evolution, swelling of a cell, or venting of a sealed cell.
Caution:
The
EMERGENCY SHUTDOWN
switch only prevents current flow from the cell
into or out of the Reference 30k Booster’s high-current cell leads. It does not protect against dangerous cell
conditions that do not involve current flow into the booster, such as short circuits within the cell.