The security testing has to be carried out in compliance with standard IEC 62353 ("Medical electrical
equipment - Recurrent test and test after repair of medical electrical equipment") or the appropriate local
derivative.
During the recurrent tests, any differences in potential between the touchable, conductive elements of
the connector assembly has to be measured. Any detected difference in potential indicates faulty secur-
ity technology, which has to be corrected. These measurements are carried out as supplementary to
the required standard tests above.
Measurement technology testing are carried out by GAMMA or by a service center authorized by
GAMMA.
The measuring flags are subjected to relatively high mechanical stresses, and undergo the results of
normal attrition. They should be replaced after approximately 500 examinations.
8.4
Operating conditions
The device is designed for operation exclusively in dry areas.
·
Temperature:
From 10 °C to 35 °C.
·
Relative humidity:
From 10% to 85%, non-condensing.
In case of a drastic change in temperature, wait until the device has reached room temperature before
beginning operation.
8.5
Storage and transport
If the device is to be transported over long distances, pack it in its original packaging. Store the flags
and styli in the appropriate containers as long as they are not in use.
For storage and transport, make sure to comply with the following environmental conditions:
·
Temperature:
From -10 °C to 60 °C.
·
Relative humidity:
From 10% to 85%. Store in a dry place!