![Gamma Vacuum DIGITEL MPCq Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/gamma-vacuum/digitel-mpcq/digitel-mpcq_instruction-manual_3377554047.webp)
Hex
command
Description
Data
field
Response Data and response interpretation
78
TSP SET
SUBLIMATION
TARGET LEVEL
Sets sublimation
target power level.
R, X, M
Where:
R is TSP (1-2)
X is number of amps/watts
M is 'A' for Amps or ‘W’ for Watts
79
TSP SET PARAMETERS
Configures TSP
parameters.
R, M, N,
X.XE-XX,
Y.YE-YY,
D, P
Where:
R is TSP (1 or 2)
M is cycle period in minutes (time
between fires)
N is number of cycles
X.XE-XX is high pressure limit
Y.YE-YY is low pressure limit
D is sublimation duration in
seconds
P is pressure window (0 = on, 1 =
off/ignore)
8B
TSP SET SUPPLY
Configures TSP
supply.
R, S
Where:
R is TSP (1-2)
S is supply (0-2), 0 indicates no
supply, pressure window is off
8C
TSP GET SUPPLY
Reads TSP supply.
R
S
Where:
R is TSP (1-2)
S is supply (1-2)
DE
TSP GET TSP STATUS
Reads TSP status.
R, X
YY
Where:
R is TSP (1-2)
X is option. Currently option 0 is
only supported.
YY is a two digit status code
‘’00” - for TSP Initializing
“01” - for TSP Off
“02” - for TSP Armed (Running)
“03” - for TSP Armed (Waiting,
pressure window, supply is off)
“04” - for TSP Armed (Waiting,
pressure window, pump pressure
outside the defined pressure
window)
“05” - for TSP Armed (Waiting,
Interlock not satisfied)
“06” - for TSP Armed (Waiting,
Other TSP is firing)
“07” - for TSP Firing
Page 47
900034_C
900034_C - Serial communication