6

  Connect the larger end of the 

Connector Hose (with the red 
button) to the Control Unit. 
Make sure you hear it click. 
To disconnect, simply press 
the red button and remove 
the connector from the 
Control Unit.

7

  Apply the selected Wrap (consult the Use Guide accompanying each Wrap for 

application instructions). To ensure proper performance of the System, it is important 
to use ATX Series Heat Exchangers in your Wrap.

Caution

  The Wrap is not sterile. Do not place directly against open wounds, sores, 

rashes, infections, or stitches. The Wrap may be applied over clothing or 
dressing.

  To ensure best fit, be sure the Wrap is completely deflated prior to each 

application.

  

If you have any problems with the set-up of your GRPro System, in the U.S. call 
Game Ready Customer Service at 1.888.426.3732 (+1.510.868.2100); from outside 
of the U.S. please contact your local distributor. 

Warnings

  Follow the treatment recommendations of your health care practitioner  

for the use of this device. 

  Improper placement or prolonged use of the GRPro could result in tissue 

damage 

  For additional warnings and precautions, please refer to pages 12 & 13  

of this manual. 

8

  Connect the smaller end of the Connector Hose (with a blue or gray button) to the 

Wrap. Make sure you hear it click. To disconnect, simply press the blue or gray button 
and remove the connector from the Wrap.

9

  Set your target temperature (displayed in the Target Temperature window) by 

turning the knob. Turning the knob clockwise, towards 3 snowflakes, will  
provide more cold therapy. Turning the knob counter-clockwise, towards 1 
snowflake, will provide less cold therapy.

4

  

Place the Control Unit in the location where you plan to use it. The Control Unit should 

only be used in the upright orientation as shown. The Control Unit will leak if placed on 
its side.

Disconnect

Button

Connector Hose

to Wrap

2

  Add water to fill-line indicated on label within the reservoir. DO NOT OVERFILL. Add 

ice to top of reservoir.

3

  Close the ice box door. Make sure you hear it click.

Click

EN-7

EN-6

Disconnect Button

Click

5

  Connect the AC Adapter to the receptacle on the end panel of the Control Unit, then 

plug the AC Adapter into a grounded electrical outlet. The power indicator light (orange 
color) on the Control Unit should illuminate. Press the power button.      The screen 
should light up and theaControl Unit should beep twice. The power indicator light 
should turn from orange to 
green. If you do not see these 
indicators, refer to “Control 
Unit will not turn on” on page 
17 of the Troubleshooting 
Guide in this manual. 

Note

: The Control Unit should 

be turned on prior to attaching 
a Wrap.

AC Adapter

to electrical outlet

SUPERSEDED

Содержание GRPro 2.1

Страница 1: ...odel Numbers 550550 03 550550 03 RN 550550 53 GRPro 2 1 Control Unit User s Manual53 Game Ready Control Unit User s Manual PN 703758 Rev D 2011 CoolSystems Inc All rights reserved Brought to you by Co...

Страница 2: ...of Operation______________________________________ 2 Buttons_________________________________________________ 2 Adjusting Temperature___________________________________ 3 Display______________________...

Страница 3: ...n on Programs Play Pause Use this button to start or pause a treatment Add Time Use this button to add time in Manual Mode does not work in Program Mode You can add up to 90 minutes Treatment must be...

Страница 4: ...a table during use Note that using the System in an environment with a high ambient temperature may affect its ability to provide adequate cooling or may limit the ice life If you will be using the S...

Страница 5: ...Make sure you hear it click To disconnect simply press the blue or gray button and remove the connector from the Wrap 9 Set your target temperature displayed in the Target Temperature window by turnin...

Страница 6: ...id your warranty Abrasive materials to clean the Control Unit Doing so will damage the plastics and will void your warranty Caution The Control Unit is not a waterproof device Do not apply a direct st...

Страница 7: ...tomer Service at 1 888 426 3732 1 510 868 2100 from outside of the U S please contact your local distributor 1 Identify the filter within the ice reservoir 2 Using two fingers grasp and squeeze the tw...

Страница 8: ...e in the acute stages of inflammatory phlebitis in the affected region Who have any history or risk factors for deep vein thrombosis or pulmonary embolus including prolonged bed rest in the affected r...

Страница 9: ...armful ingress of water 5 3 This product provides ordinary protection against ingress of water Degree of safety in the presence of flammable anesthetics or oxygen 5 5 Not suitable for use in an oxygen...

Страница 10: ...ufacturer s Declaration Electromagnetic Immunity The GRPro 2 1 is intended for use in the electromagnetic environment specified below The customer or the user of the GRPro 2 1 should assure that it is...

Страница 11: ...g NO Pressure wrap shall be vented to atmosphere The cooling temperature of the circulating ice water will be adjustable between 34 F 1 C and 50 F 10 C as long as the ice water in the ice box is suppl...

Страница 12: ...e connections Disconnect and reconnect Wrap from Connector Hose Make sure there is water in the ice box Verify that the ice box filter is not clogged Make sure there are no kinks in the Wrap or Connec...

Отзывы: