AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
215 Single needle stroke volume is too low
215 Single needle stroke volume is too low
Appears:
When the measured stroke volume is lower than the set
stroke volume in single needle treatment.
The performance of the patient's blood access may be
decreased.
The single needle treatment may not be optimised.
Machine actions:
None.
Operator action:
1.
Check the level in the expansion chamber. It should be
as low as possible but air should not pass into the blood
lines at the end of the venous phase.
2.
If possible, adjust the venous pressure alarm window.
The wider the window, the greater the stroke volume.
The higher the position of the window (i.e. how high in
the mmHg scale the window is placed), the higher the
venous flow.
3.
If previous steps do not help, adjust the stroke volume
in the dialysis machine:
- Press the
Blood
button.
- Select the
Single needle
tab.
- Press
Minimum stroke volume limit
.
- Adjust the limit value.
For more information, see Section 5
handbook.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...