HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Disinfection with the AK 98 dialysis
machine and WRO system
A:163
16)
The chemical disinfection procedure of the
WRO unit will now continue with Dwell
period followed by Rinse. The WRO unit
then goes to standby.
17)
Rinse of the dialysis machine.
●
Auto rinse, if selected, will take place
at the preset time.
●
If the dialysis machine is to be
manually rinsed this can now be
initiated at any time on the dialysis
machine.
18)
Perform a test for residual chemicals. See
Section 10.4.5
11.4
Settings for rinse
11.4.1
Rinse settings
The rinse in the integrated chemical disinfection program can either be performed
manually or automatically at a pre-defined time (Auto rinse). Auto rinse requires a
communication cable between the units. For Auto rinse, the desired ready time (day,
hour, minute) must be set in the program. Auto rinse must be activated before the
dialysis machine is switched off.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...