AK 98 Dialysis Machine - Profiling
A:119
7.3
Profiling of ultrafiltration rate
There are three available profiles for adjusting the ultrafiltration rate.
●
Linear mode
●
Step mode
●
Interval mode
Changes in the ultrafiltration rate are determined by:
●
total UF volume.
●
treatment time.
●
starting value for the
UF rate
.
●
in
Step
mode:
No of steps
.
●
in
Interval
mode:
No of intervals
.
Based on these parameters, the machine will perform an adjustment of the
UF rate
during the treatment time when profiling has been activated.
Before activating a profile for ultrafiltration rate, set the patient’s ultrafiltration volume
and treatment time.
Always check the ultrafiltration profile settings after changing ultrafiltration volume or
treatment time when a profile is active, since such changes affect the ultrafiltration
rate.
Figure 7-2. Decreasing linear mode for UF rate profiling
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...