F o r m e d i c a l
p e r s o n n e l , h e a l t h
e d u c a t i o n o n l y
)
Cleaning Instruction
Rinse thoroughly with cold
boiling water to remove
the disinfectant .
3
A i r Dry
2/ 3
1
S o a k
After removing secretions, immerse 2 / 3 of the head of
Resuscitator ( excluding the pressure relief valve) in cold
boiled
water
Repeatedly gently press the resuscitator bag to clean the
inside of it.
2
C l e a n
Reusable Resuscitator- Cleaning Instruction
Disposable Resuscitator- Cleaning Instruction
4
R i n s e T h o r o u g h l y
AD
U
LT
M
R
-
10
0
T
M
P r e s s ur e
r e li e f
v alv e
60 cm H 2 O
ADULT
MR- 100
TM
1
D i s a s s e m b l e
AD
U
LT
M
R
-
10
0
T
M
60c m H 2
O
5
Air Dry
6
Assembly
T he pressure relief v a lv e c an
be cleaned and disinfect
without being disassembled.
Clean with neutral detergent
( not recommended for oxyg e n
reser v o ir bag )
3
D i s i n f e c t i o n
AD U LT
TM
MR-100
1. Sterilize one ingot with 2 0 0 0cc
water for 30 minutes
2. Soak in 75 % alcohol for 30 minutes
3. Soak in vinegar and water ( 1 : 3
ratio ) for 30 minutes
(F o r m e d i c a l p e r s o n n e l , h e a l t h e d u c a t i o n o n l y )
© 2016 GaleMed Corp.
GM-DM-01-TW07
Assemble and test the function (Press
the patient end , press the ball , connect
the oxygen to the full pressurestate,
and the oxygen storage bag expands)
(exclude
pressure relief
valve)
Manual Resuscitator
-
Dispo-Bag Manual Resuscitator
Rescu-6
TM
Manual Resuscitator
MR-100
TM
Manual Resuscitator
Reusable
Disposable
Disposable
Press to remove excess water and leave to dry in a cool place.
GaleMed Corp.
C l e a n i n g