Introduction 

The MR‐100™ Plus    Manual Resuscitator Is designed for use as an adjunct 

to artificial respiration and cardiopulmonary resuscitation. The Resuscitator 

can be used to ventilate the apnoeic patient and to augment ventilation 
and/or oxygen delivery to the spontaneously breathing patient.   

 

When connected to an oxygen source, the MR‐100 

TM 

Plus    Manual 

Resuscitator can also deliver supplemental oxygen with the aid of an oxygen 

reservoir The oxygen concentration delivered by the MR‐100

TM

 Plus   

Manual Resuscitator depends on the oxygen supply flow rate, tidal volume, 

ventilation frequency and operator technique.   

 

The Infant and Child models of the MR‐100™ Plus    Manual Resuscitator are 

fitted with pressure relief valves      which vent the delivered breath to the 

atmosphere if the pressure exceeds 40±5cmH20. The pressure relief valve 

can be overridden by the operator should a higher delivery pressure be 

required.   

 

The MR‐100™ Plus    Manual Resuscitator is intended for use by qualified 

medical and emergency personnel trained in pulmonary ventilation and 

advanced cardiac life support techniques. Users should read and understand 

the contents of this manual and demonstrate proficiency in the assembly, 

disassembly and use of this device prior to use. 

 

 

Warning And Cautions 

 

WARNING: Do not use the MR‐100"' Plus    Manual Resuscitator In toxic 

atmospheres. 

 

WARNING: Remove the oxygen reservoir and reservoir valve If 
supplemental oxygen Is not being administered. Failure to do so will 

affect the refill rate and maximum frequency capabilities. 

 

WARNING: Do not administer supplemental oxygen in the presence of 

open flames. 

 

WARNING: Do not use oil, grease or other hydrocarbon based 

substances on any part of the MR‐100

TM 

Plus    Manual Resuscitator. 

Supplemental oxygen, 

 

supplied under pressure, can combine with hydrocarbons and cause 

explosions. 

 

WARNING: This device is intended for use by qualified medical and 

emergency personnel trained in pulmonary ventilation and advanced 

cardiac life support techniques. 

 

WARNING: Always test the device in accordance with this manual after 

cleaning and disinfection or replacement of parts. 

 

WARNING: Do not use the pressure relief valve over rid mechanism for 

the child and infant resuscitator. 

 

CAUTION: Use only replacement parts intended for use on MR‐

100™Plus    Manual Resuscitators. 

 

CAUTION: Do not attempt to disassemble the pressure relief valve 

assembly. Disassembly will damage the component. 

 

 
 
 

Содержание 2038

Страница 1: ...REF SAP REF Description 2038 AR0051 MR 100 Plus Intake Valve Kit ...

Страница 2: ...uld read and understand the contents of this manual and demonstrate proficiency in the assembly disassembly and use of this device prior to use Warning And Cautions WARNING Do not use the MR 100 Plus Manual Resuscitator In toxic atmospheres WARNING Remove the oxygen reservoir and reservoir valve If supplemental oxygen Is not being administered Failure to do so will affect the refill rate and maxim...

Страница 3: ...ent side of the duckbill valve F or when the operator releases pressure from the silicone bag 8 The valve lifts directing the patient s expiratory gases through the expiration ports G of the non rebreathing valve see figure 3 Figure 3 lnspiration The silicone bag refills for the next breath during patient exhalation Negative pressure within the bag caused by the expansion of the compressed bag ope...

Страница 4: ...will stay in the overridden position Lock Arrow forward to the NRV direction as Figure 5 1 Figure5 1 Figure 5 Overriding the Pressure Relieve Valve Operating Instruction Note lf the patient is intubated or has a tracheostomy the resuscitator can be connected by removing the mask and attaching the non rebreathing valve outlet directly to the connector of the endotracheal or tracheostomy tube 1 Posi...

Страница 5: ...g valve as follows Squeeze the silicone bag to deliver several sharp breaths through the non rebreathing valve to expel the contamination If the contamination does not clear continue ventilation with another resuscitator or use mouth to mouth mask techniques Disassemble the non rebreathing valve and rinse it with water Reassemble and disinfect Once the procedure is completed clean disinfect and te...

Страница 6: ...es are without an oxygen reservoir 02 flow In Tidal vol ml x ventilation rate L min 70x20 70X30 200X20 200X30 300X20 300X30 02 99 65 95 65 56 37 50 37 45 34 40 32 05 99 86 99 83 99 44 96 43 84 37 88 36 10 99 93 99 90 99 46 99 45 99 45 99 43 Conditions compliance 0 01L cmH2O Resistance 200 cmH2O L s l E ratio 1 2 Infant Values In parentheses are without an oxygen reservoir 02 flow In Tidal vol ml x...

Страница 7: ...ance 0 01L cmH2O Resistance 20cmH2O L s Child B VT 300ml Compliance 0 02 L cmH2O Resistance 20cmH2O L s Infant A VT 20ml Compliance 0 001 L cmH2O Resistance 400cmH2O L s Infant B VT 70ml Compliance 0 01 L cmH2O Resistance 20cmH2O L s Specifications Storage temp 40 C 10 F to 60 C 140 F Operating temp 18 C O F to 50 C 122 F Materials Silicon rubber Mask Duckbill valve Mushroom valve Relief valve sea...

Страница 8: ... rebreathing valve outlet 3 Release the bag The bag should expand immediately and refill If not check the intake valve at the base of the silicone bag is correctly assembled or not 4 While keeping the non rebreathing valve outlet blocked compress the bag again The bag should not be compressed easily If this occurs check that you are blocking the valve sufficiently and that the intake valve at the ...

Страница 9: ...ressure relief valve Oxygen Reservoir Intake Valve assemblies 1 Attach the reservoir to the intake valve inlet 2 Inflate the reservoir and block the reservoir port 3 Compress the reservoir bag Gas should escape through the safety outlet valve on the reservoir valve If not check if the reservoir valve is correctly assembled 4 Cycle the Resuscitator through several ventilations The safety inlet valv...

Страница 10: ...nd Storage Wall mount holder for resuscitators Silicone hanger for resuscitators Storage case Order information REF SAP REF Description 2038 AR0051 MR 100 Plus Intake Valve Kit 2150 AR0056 MR 100 Plus Resuscitator set adult w Pop 2151 AR0057 MR 100 Plus Resuscitator child 2152 AR0058 MR 100 Plus Resuscitator infant 2163 AR0063 MR 100 Plus Resuscitator adult ...