5.5
Troubleshoot
Soft reset
The soft reset is the first manipulation which can be done if the Physilog® is not working
properly. To do a soft reset, hold the main button for 15 seconds. The LED shines orange
and then stops. If the problem persists, fill out the contact form on the support page
(
) or write an email to: [email protected]. Please indicate
Physilog® serial number and describe the problem in as much detail as possible.
LED indications
Situation
Description
Meaning
Connected via USB
Green continuous
Fully charged
Orange blinking
Charging
Multicolor blinking
Data transfer
Red continuous
Error 1
During
standalone
measurement
Green blinking
Measuring
Green continuous
Waiting for synchronization
Orange blinking
Low battery
During measurement
started from
application
Green blinking (2 flashes per blink)
Measuring and sending data
to application
Orange blinking (2 flashes per blink)
Measuring, connection to
application to send data lost
Green/orange blinking (1 green, 1 orange flash per
blink)
Measuring and sending data
to application with low battery
Error indications
2 rapid red blinks, pause
– repeated 3 times in total Error 2
3 rapid red blinks, pause
– repeated 3 times in total Error 3
4 rapid red blinks, pause
– repeated 3 times in total Error 4
5 rapid red blinks, pause
– repeated 3 times in total Error 5
6 rapid red blinks, pause
– repeated 3 times in total Error 6
Troubleshoot for error indications
Error Type*
Troubleshoot
Error 1
Disconnect and replug the Physilog®
Error 2
Contact Gait Up
Error 3
Charge the Physilog®
Error 4
Contact Gait Up
Error 5
Contact Gait Up
Error 6
Empty the memory card
*see LED indication above