6
REV. 7/14
S/N: 1070 AND ABOVE
PLEASE CONTACT GAGE BILT FOR ALL OTHER SERIAL NUMBERS.
TROUBLE SHOOTING
Providing all maintenance conditions have been met, follow this systematic approach to diagnosis.
1. MORE THAN ONE PULL IS REQUIRED TO BREAK RIVET.
a) Tool needs to be bled. (See filling and bleeding instructions.)
b) Spring has fatigued, replace.
c) Jaws are stripped or packed with chips. Clean or replace.
d) Incorrect nose tip.
2. SLOW OR PARTIAL OPERATION WHEN THE LEVER ASS’Y (724017) IS DEPRESSED
a) Poly Seal (A-1135) and O’ring (400792) could be worn or damaged. Replace.
b) Back-up Rings (401089) and O’ring (S834) could be worn or damaged. Replace.
c) Muffler (744143) or filter inside valve spool ass’y (743142) may be plugged with dirt. Clean thoroughly and
back-blow with compressed air.
d) Hole in metering screw in valve spool ass’y (743142) may be blocked or damaged. Hole diameter should be
.028". Clear and size or replace.
3. NO OPERATION WHEN LEVER ASS’Y (724017) IS DEPRESSED
a) Tool seized due to mechanical failure or damaged parts.
4. OIL LEAKAGE
a)
DO NOT
OPERATE WITH OIL LEAKING FROM TOOL. HIGH PRESSURE OIL MAY CAUSE SEVERE PERSONAL
INJURY.
b) Any oil leaking externally should be traced to its source. An o'ring or seal that leaks should be replaced.
5. AIR BYPASS FROM VALVE HOUSING
a) If the spring (744144) breaks or dislodges, air will flow freely through the muffler (744143). Replace or reset.
Valve spring installation tool (744151) is recommended.
b) Check o'rings on valve sleeve (743144), valve spool ass’y (743142), and valve plug (744142). If worn or
damaged, replace. Valve sleeve removal tool (744152) is recommended.
6. FASTENER STEM JAMMED IN NOSE ASSEMBLY
a) Nose assembly components require service.
DISCONNECT AIR FROM TOOL
, remove the nose from the tool
and disassemble. Replace worn or broken parts. Clean the surface the jaws ride on.
b) Stems lodged side by side in the follower. Disassemble, remove stems, and reassemble.
c) Incorrect follower.
7. PINTAIL IS NOT BEING VACCUMED THROUGH.
a) Check vacuum pressure using vacuum gage 703567 (sold separately).
Note: Offset nose assemblies must be check by removing the vacuum line from the nose assembly.
b) Press gage against the vacuum line or nose assembly to create seal.
Optimum vacuum pressure at nose assembly must be between 15-22 in.Hg