12
lb Model
Capacity, division
kg Model
Capacity, division
60.000 lb
capacity : 30/60 lb
division : 0.01/0.02 lb
30.000 kg
capacity : 15/30 kg
division : 0.005/0.01 kg
150.000 lb
capacity : 60/150 lb
division : 0.02/0.05 lb
60.000 kg
capacity : 30/60 kg
division : 0.01/0.02 kg
300.000 lb
capacity : 150/300 lb
division : 0.05/0.1 lb
150.000 kg
capacity : 60/150 kg
division : 0.02/0.05 kg
500.000 lb
capacity : 250/500 lb
division : 0.1/0.2 lb
200.000 kg
capacity : 200 kg
division : 0.05/0.1 kg
4.2.2
How to Proceed with General Calibration Mode
: It adjusts calibration to display weight properly. To enter General Calibration Mode,
press “
TARE
” key ”C
ZERO
" while is shown on the display to display.
Stage
Description
Zero Adjustment
It adjusts zero point. When “C ZERO” is shown on the display, empty
the platform and press “
TARE
” Key once the weight is stabilized to
save the current zero point.
Calibration
Completed
Select a standard weight to use for calibration, Place a weight
equivalent to the weight displayed on the display on the platform and
press “
TARE
” Key to start weight adjustment.
-
UNIT
Key: Use to change the standard weight.
-
TARE
Key: Start weight adjustment.
END
Once the weight adjustment is completed, “END" is shown on the
display and Measurement Mode is started.
4.2.3
How to Proceed with Multi Calibration
: It allows to adjust weight through 3 sections. Press “UNIT” Key while “CAL” is shown on the
display to display "C ZERO" and proceed with multi calibration.
Stage
Description
Zero Adjustment
It adjusts zero point. When “
C ZERO
” is shown on the display, empty the
platform and press “
TARE
” Key once the weight is stabilized to save the
current zero point.
Weight Adjustment
for Section 1
Place a weight equivalent to the weight displayed on the display on the
platform and press “
TARE
” Key to start weight adjustment for section 1.
Weight Adjustment
for Section 2
Place a weight equivalent to the weight displayed on the display on the
platform and press “
TARE
” Key to start weight adjustment for section 2.
Weight Adjustment
for Section 3
Place a weight equivalent to the weight displayed on the display on the
platform and press “
TARE
” Key to start weight adjustment for section 3.
Calibration
Completed
Once the weight adjustment is completed, “
END
" is shown on the display
and Measurement Mode is started.
Содержание GL-6000L Series
Страница 2: ...2 ...
Страница 5: ...5 ...
Страница 6: ...6 ...
Страница 8: ...8 2 2 Product Size ...
Страница 22: ...22 9 How to Download Program 9 1 Downloader Connection ...
Страница 23: ...23 9 2 How to Download STEP1 Initiate HOPE3000FOR8051 program STEP2 Click File Open and Download ...
Страница 24: ...24 STEP3 Choose hex file STEP4 Select HT85F2280 and click 0K ...
Страница 25: ...25 STEP5 Choose ICP e C0N12C and click OK STEP6 Click Smart Programming ...
Страница 29: ...29 11 How to Sealing ...
Страница 30: ...30 12 GL6000L Series MS EXPLODED ...
Страница 31: ...31 ...
Страница 33: ...33 MEMO ...
Страница 34: ...34 MEMO ...