the symbol shows. At this point, trigger may be artificially
pressed, or the jaws have been seriously polluted, and can
no longer continue to measure.
⑵
.
“
Er
” Boot-strap error symbol, May be pressing trigger or jaw
has been opened when the power is turn on.
⑶
.
Symbol of low battery voltage: when the battery voltage
is lower than 5.2V, the symbol shows, it cannot guarantee
accuracy of the measurements. Batteries should be
replaced.
⑷
.
"
OL Ω
" symbol indicates that the measured resistance has
exceeded the upper limit of the Meter.
⑸
.
"
L0.01Ω
" symbol indicates that the measured resistance has
exceeded the lower limit of the Meter.
⑹
.
"
OL A
" symbol indicates that the measured current has
exceeded the upper limit of the Meter.
⑺
.
Alarm symbol: when the measured value is greater than
the critical value of alarm setting, the symbol flashes, and the
meter issued by intermittent "beep--beep –beep--" sound.
⑻
.
MEM
Symbol of data storage,flashing when saving data
⑼
.
MR
Review data symbol, display when viewing data, and
display the number of stored data at the same time.
⑽
.
NOISE
symbol: This symbol flashes when there is a large
interference current in the ground loop being tested. At the
same time, the instrument sounds a
“
beep-beep-beep-
”
prompt. At this point, the accuracy of the test cannot be
guaranteed.