D-15
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the HealthGuard-15
The HealthGuard-15 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the HealthGuard-15 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the HealthGuard-15 as recommended below,
according to the maximum output power of the communications equipment.
Separation Distance According to Frequency of Transmitter
m
Rated Maximum
Output Power of
Transmitter
W
150 kHz to 80 MHz
d = 1.2
√
P
80 MHz to 800 MHz
d = 1.2
√
P
800 MHz to 2.5GHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.