SAFETY

Use extreme caution when the laser beam is turned on.

Do not let the beam enter your eyes, another person’s eyes or the eyes of an an-

imal. Be careful that reflections of the beam (on a reflective surface) do not strike 

your eyes.

Do not aim the laser beam at any gas that may explode.

Please read the complete safety instructions in the booklet delivered with this 

device.

HOW IT WORKS

Infrared thermometers measure the surface temperature of an object. The unit’s 

optics sense emitted, reflected and transmitted energy, which is collected and 

focused onto a detector. The unit’s electronics translate the information into a 

temperature reading, which is displayed on the display. The laser in this device is 

used for aiming purposes only.

FIELD OF VIEW

Make sure the target is larger than the unit’s spot size. The smaller the target, the 

closer you should go to make your measurement. When accuracy is critical, make 

sure the target is at least twice as large as the spot size.

DISTANCE & SPOT SIZE

As the distance (D) from the object increases, the spot size (S) of the area meas-

ured by the Temppointer 9 becomes larger. 

The relationship between distance and spot size for this unit is 16:1. This means 

that when measuring at a distance of 16cm of the object, the measuring surface 

has approximately a diameter of 1cm. 

MEASUREMENT OPERATION

 · Hold the meter by its handle grip and point it towards the surface to be meas-

ured.

 · Pull and hold the trigger (10) to turn the meter on and begin testing. The SCAN 

symbol (A) will appear on the LCD display. As long as the trigger isn’t released, 

temperatures will be scanned.  

To locate a hot-spot: Aim the thermometer outside the area of interest. Scan 

across with an up and down motion until you locate a hot spot. 

 · If there appears nothing on the screen, please check the batteries. 

 · Release the trigger and the HOLD symbol (B) will appear on the LCD display. 

This indicates the last temperature measurements stays available in the device. 

 · The meter will automatically power down approximately 10 seconds after the 

trigger is released.

NOTE

 

·

An infrAred thermometer must AdApt to Ambient temperAture. this mAy tAke 

up to 30 minutes to Adjust to high temperAture fluctuAtions. so, wAit A few 

minutes between meAsuring A hot And cold object.

 

·

An infrAred thermometer cAnnot be meAsured through trAnspArent surfAces 

(eg. glAss). he will then meAsure the temperAture of the glAss.

 

·

steAm, dust, smoke ... mAke meAsurements with An infrAred thermometer less 

AccurAte to unreliAble.

EXTRA FUNCTIONS

 · As long as the HOLD symbol (B) is visible on the screen, pushing the Laser but-

ton / Up button (4) will switch on or off the laser indication. 

 

 

 

 

768 mm

384 mm

192 mm

Ø12 mm

Ø24 mm

Ø48 mm

LASER OFF

LASER ON