DISTY40 / DISTY80

11

MANUAL

Appendix

Message codes

code

reason

solution

204

calculation error

perform the measurement again

208

received signal too weak, too long 

measuring time, distance too large

use a target plate

209

received signal too strong

reflection of the goal is too great. use a target image

252

temperature too high

let the instrument cool down

253

temperature too low

warm up the instrument

255

hardware error

switch the device on and off several times. if the message conti-

nues to appear, the device may be defected. contact your dealer.

Measuring conditions

Measuring range

The Disty40 is limited to 40m, the Disty80 to 80m. At night or dusk and if the target is in shadow, the measuring range without 

target plate is increased. Use a target plate to increase the measurement range during daylight or if the target has poor reflection 

properties.

Target surfaces

Measuring errors can occur when measuring toward colourless liquids (e.g. water) or dust free glass, Styrofoam or similar semi-

perme-able surfaces. Aiming at high gloss surfaces may deflect the laser beam and lead to measurement errors. Against non-re-

flective and dark surfaces the measuring time may increase.

Care

Do not immerse the instrument in water. Wipe off dirt with a damp, soft cloth. Do not use aggresive cleaning agents or solutions. 

Handle the instrument as you would a telescope or camera.

Содержание DISTY40

Страница 1: ...DISTY40 DISTY80 201 40 201 80 MANUAL...

Страница 2: ...evice dur ing transportation Regularly inspect the accuracy of the instrument especially when starting any major square setting work You have sole responsibility for the accuracy of your work Do not u...

Страница 3: ...Off button 12 Laser active 13 Referencepoint in front 14 Referencepoint at the back 15 Reference point positioning bracket 16 Max and Min measurement 17 Surface Volume 18 Indirect measurement Single p...

Страница 4: ...n be adapted for the following measuring situations For measurements from an edge fold out the positioning bracket until it first locks in place corner of 90 For measurements from a corner open the po...

Страница 5: ...each single measurements can be deleted and remeasured in series Reference The default reference setting is from the rear of the instument The display will show you Press the button to change the refe...

Страница 6: ...ton and hold down this button until you hear a beep Then slowly sweep the laser back and forth and up and down over the desired target point e g into the corner of a room Press the button to stop cont...

Страница 7: ...rs in the display Press the button to take the first measurement e g length Press the button to take the second measurement e g widht Press the button to take the third measurement e g height The resu...

Страница 8: ...t All target points must be in a horizontal or vertical plane The best results are achieved when the instrument is rotated about a fixed point e g with the position bracket fully folded out and the in...

Страница 9: ...al sode on the second link The calculated height on the third line Keep in mind that the lenght ofthe horizontal line is only 100 correct when a precise vertical measurement is performed Double indire...

Страница 10: ...d third 90 measurement The results are now shown on the screen from top to bottom distance to first upper point distance to the second point distance from the measured horizontal point distance betwee...

Страница 11: ...ge The Disty40 is limited to 40m the Disty80 to 80m At night or dusk and if the target is in shadow the measuring range without target plate is increased Use a target plate to increase the measurement...

Страница 12: ...to counteract them The product is not to be operated until the user has been instructed on how to work with it LIMITS OF USE Environment Suitable for use in an atmosphere appro priate for permanent h...

Страница 13: ...cted to mechanical shock for example blows or falling the product may be damaged or peo ple may sustain injury When setting up the product make sure that the accessories are correctly adapted fitted s...

Страница 14: ...hazards The per son responsible for the product must ensure that all users understand these directions and adhere to them Products classified as laser class 1 class 2 and class 3R do not require lase...

Страница 15: ...lly select ed components and crafted with the utmost care By that we can offer you an optimal quality If your device should however be defective due to faulty materials or a production error you can b...

Страница 16: ...futechtools futech tools Facebook LinkedIn JOIN US futech tools com World Wide Web...

Отзывы: