SweatStop® DE20
Warning
1. The patient is the responsible user of this device and should only operate it under the instruction
or guidance of a medical professional.
2. The target population is over 18 years old and was given a medical prescription to use the
device.
3. Do not disassemble the device. If needed, return it for examination or repair.
4. CAUTION: All modifications of this equipment or its accessories are forbidden.
5. The medical device should not be used close to or on top of other electric equipment. Should
such a usage be strictly necessary, the medical device must be verified to conduct normal opera-
tion in the configuration in which it will be used.
6. CAUTION: The usage of accessories, chargers or cables other than those specified or provided by
the manufacturer of this equipment may negatively influence the electromagnetic emissions or
electromagnetic immunity of this equipment and result in a faulty operation.
7. CAUTION: Portable RF communications equipment (including peripherals such as antenna cables
and external antennae) should be used no closer than 30 cm (12 inches) to any part of the
medical equipment, including the cables specified by the manufacturer. Otherwise, the function
of the equipment might deteriorate.
8. Do not use the iontophoresis device near the active high frequency equipment and the RF-
shielded room of an MRI scanner, where the intensity of disturbances is extraordinarily high.
9. Use only with tap water. Liquids other than tap water are not allowed.
Safety measures
1. Keep away from children.
2. This device should only be used with the leads and cables provided by the manufacturer to
ensure the maximum clinical effectiveness. Using accessories which are not approved by the
manufacturer may cause damage or an unauthorised interference emission and loss of opera-
ting license.
3. To prevent burns during the treatment, make sure the electrodes are covered at all times. Direct
contact with the surface of the electrodes must be avoided at all times.
4. Any damage to the skin (minor injuries, scratches etc.) in the treatment area must be coated
with Vaseline or a fatty lotion before treatment to avoid further damages.
5. Two devices may not be used simultaneously on one patient.
6. Make sure that the device is at room temperature before operating it. If the device has been
stored in low temperatures, a condensation of the air humidity may lead to malfunctions or
damage of the device.
7. All jewellery worn in the treatment areas must be removed before starting the therapy. Due to
their increased current density, metal objects can cause local skin injuries.