Functional Cosmetics Company SweatStop DE20 Скачать руководство пользователя страница 4

4

EN

SweatStop® DE20

Warning

1.  The patient is the responsible user of this device and should only operate it under the instruction 

or guidance of a medical professional. 

2.  The target population is over 18 years old and was given a medical prescription to use the  

device.

3. Do not disassemble the device. If needed, return it for examination or repair.

4. CAUTION: All modifications of this equipment or its accessories are forbidden.

5.  The medical device should not be used close to or on top of other electric equipment. Should 

such a usage be strictly necessary, the medical device must be verified to conduct normal opera-

tion in the configuration in which it will be used.

6.  CAUTION: The usage of accessories, chargers or cables other than those specified or provided by 

the manufacturer of this equipment may negatively influence the electromagnetic emissions or 

electromagnetic immunity of this equipment and result in a faulty operation.

7.  CAUTION: Portable RF communications equipment (including peripherals such as antenna cables  

and external antennae) should be used no closer than 30 cm (12 inches) to any part of the  

medical equipment, including the cables specified by the manufacturer. Otherwise, the function 

of the equipment might deteriorate. 

8.  Do not use the iontophoresis device near the active high frequency equipment and the RF- 

shielded room of an MRI scanner, where the intensity of disturbances is extraordinarily high.

9. Use only with tap water. Liquids other than tap water are not allowed.

Safety measures

1. Keep away from children. 

2.  This device should only be used with the leads and cables provided by the manufacturer to 

ensure the maximum clinical effectiveness. Using accessories which are not approved by the  

manufacturer may cause damage or an unauthorised interference emission and loss of opera-

ting license. 

3.  To prevent burns during the treatment, make sure the electrodes are covered at all times. Direct 

contact with the surface of the electrodes must be avoided at all times.

4.  Any damage to the skin (minor injuries, scratches etc.) in the treatment area must be coated 

with Vaseline or a fatty lotion before treatment to avoid further damages.

5. Two devices may not be used simultaneously on one patient.

6.  Make sure that the device is at room temperature before operating it. If the device has been 

stored in low temperatures, a condensation of the air humidity may lead to malfunctions or 

damage of the device.

7.  All jewellery worn in the treatment areas must be removed before starting the therapy. Due to 

their increased current density, metal objects can cause local skin injuries. 

Содержание SweatStop DE20

Страница 1: ...Tabwater Iontophoresis Device SweatStop DE20 EN Instruction Manual...

Страница 2: ...cation 11 Basics 11 Preparation 12 Hand treatment 13 Foot treatment 15 Combined treatment of hand and feet 16 Underarm treatment 18 Treatment interval 19 Power supply 20 Monitored battery accumulator...

Страница 3: ...nt is the intended operator of the iontophoresis device under the instruction or guidance of the clinician The DE20 is intended to be used in medical facilities and in private homes by healthcare prof...

Страница 4: ...function of the equipment might deteriorate 8 Do not use the iontophoresis device near the active high frequency equipment and the RF shielded room of an MRI scanner where the intensity of disturbance...

Страница 5: ...plugged in firmly 14 The mains plug is specified as a part of this medical equipment Only use the adaptor supplied by the manufacturer 15 Do not submerge the device in water 16 Dropping the device ma...

Страница 6: ...arger Charger for series of 4 NiMH accumulators Order number A 0011 Cables for electrodes Order nomber Z 0066 Tubs for tap water Order number Z 3000 Electrodes in sponges for the treatment of hands an...

Страница 7: ...ller 2 Diverter for type of current 3 Outlet sockets for electrodes 4 LED indicator weak battery 5 Digital display 6 LED indicator empty battery 7 Charging socket 8 Battery compartment 9 Code control...

Страница 8: ...in the cables according to their colours to connect the device with the electrodes Change the polarity each time you start the therapy unless otherwise recommended by your doctor The pole reversal can...

Страница 9: ...apacity of the batteries and the charger new cells standard equipment 4x 2500 mAh take approximately 11 12 hours to charge 9 Code control The number of all completed treatments will be saved This code...

Страница 10: ...he real current value Disconnecting the device from the current supply 240V 110V When the DC plug of the charger is plugged into the charging socket of the device the treatment cannot be conducted The...

Страница 11: ...atment Since the causes of hyperhidrosis differ between patients you will have to adjust the therapy to your individual needs and lifestyle These factors may vary according to yearly seasons and every...

Страница 12: ...vel surface For the treatment of the hands place both trays side by side on a table for the treatment of the feet place both trays side by side on the floor If you treat hands and feet simultaneously...

Страница 13: ...your hands out of the tub at any time to readjust the amperage Please note that you should only increase the amperage slowly The increase of the current strength is restricted which means that turnin...

Страница 14: ...up the amperage until the current is noticeable then turn it down again until you can no longer feel it Pulsed current Turn up the amperage until the current is barely noticeable Iontophoresis treatme...

Страница 15: ...on the display for 5 seconds this indicates that the device is running Now check whether you have selected the correct current type The first line on the display shows if the device is set to DC dire...

Страница 16: ...hands or feet If necessary adjust the water level The current flow will be activated as soon as the electric circuit is closed The current strength will increase until the desired value has been reac...

Страница 17: ...line on the display shows if the device is set to DC direct current or PC pulsed current When turning the button 1 clockwise the amperage will be increased The second line on the display 5 shows the c...

Страница 18: ...ts prefer the comfortable treatment feeling of the pulsed current To turn on the device turn the switch 1 to the left A text will appear on the display for 5 seconds this indicates that the device is...

Страница 19: ...aused treatment every third day two times a week with a minimum pause of two days The duration is fixed to 20 minutes to allow an evaluation of the effectiveness of the treatment this leads to a reduc...

Страница 20: ...ue to this the device can be operated independently from a power connection Monitored battery accumulator capacity As soon as the battery voltage drops below a defined value the yellow LED 4 will ligh...

Страница 21: ...V peak Notice When the DC plug of the charger is plugged into the charging socket the treatment cannot be conducted The DC plug disconnects the electric circuit of the power supply Before using the de...

Страница 22: ...eries out For short term storage please check or renew the batteries periodically to keep the equipment in a fully usable condition Disposal The model DE20 iontophoresis device contains electronic com...

Страница 23: ...n Read operator manual Consult instruction for use Manufacturer Manufacture date Batch code IP21 Protected against solid objects over 12 5 mm Protected against vertically falling drops of water 2460 C...

Страница 24: ...device is suitable for use in all establish ments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domesti...

Страница 25: ...floors are covered with synthetic material the relative humidity should be at least 30 Power frequency 50 or 60 Hz magnetic field IEC 61000 4 8 30 A m 50 or 60 Hz 30 A m 50 or 60 Hz Power frequency ma...

Страница 26: ...ital environ ment If the user of the device requires continued operation during power mains interruptions it is re commended that the device be powered from an uninterrupti ble power supply or a batte...

Страница 27: ...mitter in watts W according to the trans mitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF trans mitters as determined by an electromagnetic si...

Страница 28: ...Cosmetics Company AG Nauenstr 67 4052 Basel Switzerland Distribution Center EU Schwarzwaldstr 31 79539 L rrach Germany Tel 41 61 262 10 00 Fax 41 61 303 80 09 contact functional cosmetics com www swea...

Отзывы: