3
(6) External Equipment Connector 2
Connects the HR-800 Recorder Unit via the unit connection
cable.
(7) External Equipment Connector 1
Connects to the DS-8500 System via the unit connection cable.
Connection Procedures
For more detail about how to connect a respiration circuit and the
calibration procedure, refer to the operation manual of the patient
monitor, the DS-8500 System.
The following connection cables are available depending on the
installation.
Model Type
Length
CJO-09SS0.3 0.3m
CJO-09SS1.5 1.5m
CJO-09SS5 5m
WARNING
Use only the cables specified by Fukuda Denshi.
Not only the MGU-810 cannot deliver its maximum performance
but may also result in increase in emission or decrease in
immunity.
CAUTION
Make sure that the power of the DSC-8500 Main Unit is turned
OFF when connecting/disconnecting the connection cable.
When connecting with the unit connection cable, make sure to
secure the connector with screws.
CAUTION
Precautions about the Operating Environment
The following environmental conditions should be observed
when operating the equipment.
Ambient Temperature: 10 to 35
C
Relative Humidity :30 to 85% (non-condensing)
The power is supplied from the patient monitor. Read the
patient monitor’s operation manual and then connect the
connection cable properly.
Pay attention when installing or storing the equipment. Do not
install or store in the following locations.
where chemicals are stored or gas may evolved
where the equipment will be subject to splashing water or
humidity from a nebulizer or vapolizer
where the equipment will be subject to direct sunlight
unstable place with inclination, vibration, or shock
How to Connect the DSC-8500 series Main Unit and
the MGU-810
1
Connect the main unit (DSC-8500 series) and MGU-810 using
the unit connection cable (CJO-09SSxx).
How to Connect the DSC-8500 series Main Unit
with the MGU-810 and HR-800
1 Connect the DSC-8500 series Main Unit and the MGU-810
with the unit connection cable (CJO-09SSxx).
2 Connect the MGU-810 and the HR-800 with the unit
connection cable (CJO-09SSxx).
Settings on the Patient Monitor
To use the MGU-810, the setting, such as “External Device
Connection Setup under Initial Settings”, on the DS-8500 System is
required.
For more details, refer to the operation manual of the patient
monitor, the DS-8500 System.
Measurement Accuracy
In order to perform accurate measurement, follow the precautions
below.
CAUTION
Before measurement, make sure to warm up the unit for more
than 10 minutes.
Make sure to secure the connection between this unit and the
sampling line, or the airway adaptor so that the gas will not leak.
Accuracy Check of Gas Measurement
In order to maintain the measurement accuracy, make sure to
perform the accuracy check of gas measurement once a year.
When performing the accuracy check, follow the precautions
below.
CAUTION
Before the accuracy check, make sure to warm up the unit for
more than 10 minutes.
Make sure to use the specified calibration gas which is within
the expiration date. The measurement accuracy will decrease if
a low pressure gas is used.
Manual Zero Calibration
On the patient monitor, a zero calibration for the multigas
measurement is periodically performed, but it can also be
performed manually when necessary.
For the measurement range and accuracy of each parameter, refer
to the “Specification” section below.
HR-800