Warning Label 

Make sure to read the warning label attached to the equipment and 
comply with the requirements while operating the equipment. 

CAUTION

 

  Do not damage or erase the warning label attached to the 

equipment. 

  The warning label contains important descriptions for handling 

and operating the equipment properly and safely. A damaged 
label may compromise safe operation. 

 

MGU-811P 

 

 

 

Graphic Symbols 

The following are the symbols and their meaning indicated on the 
equipment. 
 

Symbol Description 

 

Caution; refer to accompanying documents 
Indicates the need to refer to related accompanying 

documents before operation. 

 

Type BF Applied Part with Defibrillation-Proof 

Indicates the degree of protection against electric 
shock which is Type BF Applied Part with 
defibrillation-proof. 

 

 

GAS Output 

Indicates the position where it exhausts the 
sampling gas. 

 

Year of Manufacture 
Indicates the manufactured year. 

 

WEEE (Waste Electrical and Electronics Equipment)
Indicates a separate collection for electrical and 
electronic equipment. 

 

General Description

 

The MGU-810 series is a sidestream gas measuring unit. The 
following model types are available according to the different 
measurement parameters. 
 

Model Type  CO

2

/N

2

Agent GAS 

Measurement 

O

Measurement 

Spirometry

MGU-811P Yes 

Yes 

Yes* 

Yes 

MGU-812 Yes 

Yes 

No 

Yes 

MGU-813 

Yes No  No Yes 

* with the Servomex Paramagnetic Oxygen Sensor 

 
This equipment is used for the specified patient monitor. The 
contents to be displayed differ depending on the connected patient 
monitor, thus refer to the patient monitor’s operation manual, the 
DS-8500 System. 
 

Features 

 

Automatically detects the type of anesthetic gas for monitoring. 

 

With the sidestream method, the patient connection part is 
smaller, lighter and quieter. 

 

CO

2

, N

2

O, five anesthetic agents (Halothane, Isoflurane, 

Sevoflurane, Enflurane, and Desflurane), and O

2

 concentration 

can be measured. 

 

Model types with/without the Agent GAS measurement and/or 
O

2

 measurement is available. 

 

O

2

 measurement uses a paramagnetic method. 

 

Spirometry with the flow sensor allows measuring the Airway 
Pressure, Tidal Volume, Airway Flow, Minute Volume, Airway 
Compliance, Airway Resistance, and I:E ratio. 

 

Name of Parts and Their Functions

 

 

WARNING

 

Do not connect a unit or cable not authorized by Fukuda Denshi to 
any I/O connector. If done so by mistake, the equipment cannot 
deliver its maximum performance and the connected units may be 
damaged, resulting in a safety hazard. 

 

Front Side 

 

 

Rear Side 

 

 

(1) Power Supply Indicator 

Indicates the power status. 

  Green    : Power is supplied to the MGU-810. 
  Light Off : When the power of the DS-8500 System, which is 

connected to the MGU-810, is OFF, or the power supply 
indicator on the display unit is orange (in standby mode).   

 
(2) Inlet 

Connects the gas sampling line of the SPRIT

 Flow Sensor. 

 
(3) DRYLINE™ Water Trap (with container)   

Removes water inside the sampling tube connected to the patient. 
Empty the water trap container when it is half full. When 
connecting to a new patient, clean it with the specified disinfect 
solution (refer to the “Cleaning” below). When removing the 
water trap, press the side buttons. 

 
(4) Exhaust Outlet 

Connects gas exhaust system and exhausts sampling gas. 

 

 

(5) Flow Sensor Connector 

Connects the pressure line of the SPRIT

 Flow Sensor. 

 

(1)
 
(2)
 
(3)
 
(4)

(6)
 
(7)

(5)

Содержание MGU-810 Series

Страница 1: ...orrectly operate the equipment Users should have a thorough knowledge of the operation before using this equipment Pay attention to the following when installing and storing the equipment Do not install or store in an area where the equipment will be subject to splashing water Do not install or store in an area where the environmental conditions such as atmospheric pressure temperature humidity ve...

Страница 2: ...the patient monitor s operation manual the DS 8500 System Features Automatically detects the type of anesthetic gas for monitoring With the sidestream method the patient connection part is smaller lighter and quieter CO2 N2O five anesthetic agents Halothane Isoflurane Sevoflurane Enflurane and Desflurane and O2 concentration can be measured Model types with without the Agent GAS measurement and or...

Страница 3: ...direct sunlight unstable place with inclination vibration or shock How to Connect the DSC 8500 series Main Unit and the MGU 810 1 Connect the main unit DSC 8500 series and MGU 810 using the unit connection cable CJO 09SSxx How to Connect the DSC 8500 series Main Unit with the MGU 810 and HR 800 1 Connect the DSC 8500 series Main Unit and the MGU 810 with the unit connection cable CJO 09SSxx 2 Conn...

Страница 4: ...izing the entire room using a spray solution pay close attention not to have liquids get into the equipment or connectors Use only neutral detergent to clean the housing The surface resin coating may be damaged resulting in discoloration scratches and other problems Do not use chemical cloth scrub brush abrasive polishing powder hot water volatile solvent and chemicals cleanser thinner benzine ben...

Страница 5: ...quipment Degree of protection against ingress of water IPX0 no protection Protection against ignition of flammable gas Not provided Power Supply Voltage DC 18V Supplied from the DSC 8500 series Main Unit Usable Life 6 years According to self certification Refer to the above Maintenance Periodic Replacement for parts requiring periodic replacement Sampling Flow and Accuracy Flow Range 70 to 200mL m...

Страница 6: ...urement Range 2 to 20 L min Adult 0 5 to 5 L min Pediatric Compliance mL cmH2O Measurement Range 4 to 100 m cmH2O Adult 1 to 100 m cmH2O Pediatric Airway Resistance cmH2O L s Measurement Range 0 to 40 cmH2O L s Adult Pediatric Peak Plateau PEEP and Mean Pressure cmH2O Measurement Range 20 to 100 cmH2O Adult Pediatric I E Ratio Measurement Range 1 4 5 to 2 1 Adult Pediatric Conditions of use for St...

Отзывы: